14 . Peltier Temperature Programmer User's Guide

Electrical Safety

Connect the PTP to a power supply line that includes a switch or other adequate means of
disconnection from the electricity supply.
Only plug the PTP into an electricity supply socket that is provided with a protective ground
(earth) connection.
To ensure satisfactory and safe operation of the PTP it is essential that the green/yellow lead
of the line power cable is connected to a true electrical ground.
If any part of the PTP is not installed by a PerkinElmer Service Engineer, ensure that the line
power plug is correctly wired.
Terminal
Live
Neutral
Earth/Ground
Any interruption of the protective ground (earth) conductor inside or
outside the instrument or disconnection of the protective ground
(earth) terminal can make the instrument dangerous.
WARNING
Intentional interruption is prohibited.
Even with the power switch off, line power voltages can still be present
within the instrument.
Cable Lead Colors
International
Brown
Blue
Green/Yellow
USA
Black
White
Green