Control Panel: Adjusting Retention Force

Step 1: Preperation
Tool required:
Flat-head screwdriver
Allen key SW2
Step 3: Adjusting retention force
3.
Alternate between tightening the two screws until
the desired retention force is acheived.
4.
Swivel the control panel between the horizontal
position (folded down) and 100° position (folded
up) and check if the control panel maintains its
position. If the control panel moves repeat step 3
until control panel remains stationary.
Attention: Ensure that no wiring is damaged
during adjustment.
Step 2: Remove cap cover
1. Carefully place the flat-head screwdriver under the
cover cap. This is accessible by the recess between the
cap cover and the nut.
2. Gently pull the flat-head screwdriver away from the
device until the cover cap is pushed out.
Step 4: Replace the cover cap
5. Carefully reinsert the cover cap until both latching
clamps engage.