Siemens SITRANS LC500 Operating Instructions Manual page 67

Capacitance transmitters
Table of Contents

Advertisement

The transmitter variable on which this selection is based is chosen in Menu 01. The
increment/decrement step size is set at Menu 09.
Menu
Affected
Item
by:
0D
01, 07, 09
Lower Range Value
Lower Range Value (LRV) is the setting for 0% of the operating range, in most cases an
empty vessel/tank. If the display mode is in % this selection is disabled and the LCD
displays - - - -.
The transmitter variable on which this menu selection operates is chosen in Menu 01. For
TV0 the LRV is in most cases Factory Set to the probe capacitance in free air.
The increment/decrement step size is subject to the setting from Menu 09.
Menu
Affected
Item
by:
0E
01, 09, 0B, 0C E
1. Set the rotary switch to E.
2. Press and hold both buttons simultaneously to take the current PV reading as
the new setting.
or: Press the RED (+) or BLUE (–) button to step the value up or down.
or: Press and hold a button for a prolonged time to start a repeat function.
When the new setting exceeds that of the Limit Settings (Menu 0C and 0B), the new value
is rejected and the previous value remains unchanged.
7ML19985GE04
Rotary
Left
Switch
Arrow
Position
D
Off
Rotary
Left
Switch
Arrow
Position
Off
SITRANS LC500 – OPERATING INSTRUCTIONS
Mode
Description
Delta Range Setting
2-state
(enabled)
Analog
*
Mode Description Values
Lower Range
Value
Analog
Factory setting
Display percent Display shows
Values
Span
setting
Display
shows
- - - -
Range: 0.00 to 3300
Probe capacitance in
air
- - - -
Page 59