4 Installation | 4.11 Assembling the Transmitted Light Light Source

4.11 Assembling the Transmitted Light Light Source

4.11.1 Assembling the LED10 Light Source for Transmitted Light Illumination

Hazardous light emission
The light source belongs to Risk Group 2 as specified in IEC 62471 and emits LED radiation and
UV radiation. Eye damage or skin irritation may result from exposure.
4
4
4
Fig. 45: Assembling the LED10 light source for transmitted light illumination
1
3
5
Parts and Tools
Prerequisite
ü
ü
Procedure
1. At the illumination mount on the rear side of the stand, loosen the clamping screw
2. Insert the plug
3. Insert the light source's dovetail ring into the illumination mount
4. Ensure that the illumination cable is not being pinched or clamped.
5. Tighten the clamping screw
Proceed in the reverse order for removal.
74
CAUTION
Never look directly into the light-emitting aperture of the light source.
Avoid exposure of skin to the radiation. Use suitable protective equipment/protective
clothing if required.
Before installing or removing the light source always make sure it is switched off.
1
2
3 mm hex key
LED10 light source
Socket for transmitted light illumination
Hex Key, 3.0 mm
The microscope is switched off.
The stand is equipped with the illumination mount.
4
of the light source's cable into the socket
lumination.
Instruction Manual ZEISS Axioscope 5, Axioscope 5/7 MAT | en-US | Rev. 13 | 430035-7344-001
3
2
Clamping screw
4
Plug
6
Illumination mount
2
.
6
2
4
5
for transmitted light il-
6
.
ZEISS
5
2
.