Calibration Adjust Procedure - Fluke 1773 Calibration Manual

3 phase power quality analyzer
Hide thumbs Also See for 1773:
Table of Contents

Advertisement

To verify:
1. Connect the 4-pin connectors AUX 1 + (pin 1) to the 8846A INPUT HI.
2. Connect the 4-pin connectors AUX 1 – (pin 2) to the 8846A INPUT LO.
3. Connect the 17xx AUX Adapter box AUX 1 + and AUX 2 + to the calibrator Normal HI 4.
4. Connect the 17xx AUX Adapter box AUX 1 – and AUX 2 – to the calibrator Normal LO.
5. Set the 8588A to DC V.
6. Apply the voltages in
7. Verify that the AUX 1 readings are between the limits.
8. After the values are checked for AUX 1, move the 4-pin connectors leads to AUX 2; Pin 3 to
the 8588A INPUT HI; pin 4 connected to the 8588A INPUT LO.
9. Apply the voltages in
Calibrator Out DC Volts Vdc
300.000
600.000
990.000
10. When finished, set the calibrator to Standby.

Calibration Adjust Procedure

This procedure adjusts the Analyzer accuracy at ambient temperature 23 °C ±5 °C (intrinsic
error). The required equipment and cables for calibrating the Product are listed in
USB Communication
To prevent electrical shock, personal injury, or fire:
Do not perform the calibration procedures or calibration verification tests
described in this manual unless you are qualified to do so.
Repairs or service should be performed only by qualified personnel.
The Calibration Tool contains an automated adjust. When used, it:
provides connection instructions
can control the calibrator to apply the required voltage
will query the measurement readings of the DMM
will calculate and store the new calibration factors
Table
17.
Table
17. Verify that the AUX 2 readings are between the limits.
Table 17. AUX Input Verification
2.9848
5.9698
9.8503
for instructions on how to set up the PC.
XW Warning
Lower Limit Vdc
3 Phase Power Quality Analyzer
Calibration Adjust Procedure
Upper Limit Vdc
3.0152
6.0302
9.9497
Table
2. See
23