Calibration Adjust Procedure

This procedure adjusts the 1730 accuracy at ambient temperature 23 °C ±2 K
(intrinsic error).
The required equipment and cables for calibrating the Product are listed in
Table 2. See USB Communication for instructions on how to set up the PC.
To avoid electrical shock, personal injury, or fire:
• Do not perform the calibration procedures or calibration
verification tests described in this manual unless you are
qualified to do so.
• Repairs or servicing should be performed only by qualified
personnel.
The spreadsheet contains an automated adjust in the Calibration & Verification
worksheet. When used, it provides connection instructions, can control the
calibrator to apply the required voltage, and then will calculate and store the new
calibration factors.
When this worksheet is active, the selection box should pop up. If not, click Start
button on the upper right of the worksheet.
In the Setup tab, only the Calibrator control needs to be set (the Voltage divider
is not used in the 1730 Adjust).
Select the Calibration tab of the 1730 Calibration & Verification pop-up. Choose
Voltage, AUX, or Current calibration and check the boxes to select items for
adjustment.
1. When selection is complete, click Start.
2. Follow the instructions provided in the automated procedure.
When the 1730 Calibration & Verification popup box shows again, the calibration
factors have been calculated and stored in the 1730.
This concludes the calibration.