Rinstrum Pty Ltd. Disclaimer Rinstrum Pty Ltd reserves the right to make changes to the products contained in this manual in order to improve design, performance or reliability. The information in this manual is believed to be accurate in all respects at the time of publication, but is subject to change without notice.
Page 3
Quick Start Manual Rev 2.45 - Software Versions 3.xx GROUP ITEM SELECT EDIT (GRP) (ITM) (SEL) (EDT) Underline = Defaults BUILD Decimal Point Position 000000, 00000.0, 0000.00, 000.000, 00.0000, 0.00000 Accept DP⊗ Maximum Capacity 000100 to 999999 Default = 003000 Accept CAP⊗...
Page 4
Quick Start Manual Rev 2.45 - Software Versions 3.xx GROUP ITEM SELECT EDIT (GRP) (ITM) (SEL) (EDT) Underline = Defaults OPTION Enable Rear Pushbutton OFF, ON Accept R.ENTRY⊗ (continued) for Full Setup Entry Zero Calibration Routine <SEL> key to start. Current weight displays. ZERO⊗...
Page 5
Quick Start Manual Rev 2.45 - Software Versions 3.xx GROUP ITEM SELECT EDIT (GRP) (ITM) (SEL) (EDT) Underline = Defaults Edit Linearisation Points <SEL> to step through list of points. ED.LIN⊗ Set between zero and full <OK> to change selected point or <ITM> to exit. (continued) (K302 only) scale...
Page 7
Quick Start Manual Rev 2.45 - Software Versions 3.xx GROUP ITEM SELECT EDIT (GRP) (ITM) (SEL) (EDT) Underline = Defaults SPEC AUT.OFF Auto Power Off / Battery NEVER, 1, 5, 10 (time in minutes) Accept Operation Default = Never powers off automatically (continued) (Battery Mode: powers down after 30 minutes) B.LIGHT...
Page 8
Quick Start Manual Rev 2.45 - Software Versions 3.xx GROUP ITEM SELECT EDIT (GRP) (ITM) (SEL) (EDT) Underline = Defaults SERIAL BITS Serial Format Options N 8 1 - (Default Serial Format Options) Accept (continued) Position 1: Parity <SEL> changes position, <EDT> changes digit. Position 2: Data Bits Parity: N None, O Odd, E Even Position 3: Stop Bits...
Page 9
Quick Start Manual Rev 2.45 - Software Versions 3.xx GROUP ITEM SELECT EDIT (GRP) (ITM) (SEL) (EDT) Underline = Defaults SET.PTS TARG.2 Target for Over and Under -99999 to 999999 Accept (continued) Weight Default = 000000 Setpoint 2 (Output 2) CLOC FORMAT Date Format Setting...
Page 10
Quick Start Manual Rev 2.45 - Software Versions 3.xx GROUP ITEM SELECT EDIT (GRP) (ITM) (SEL) (EDT) Underline = Defaults TEST FRC.OUT Force Outputs <EDT> advances through outputs (continued) (ie. ON.1 and ON.2) <OK> turns outputs off and exits test. Default = OFF O.LOAD Input Overload Count...
Page 11
Quick Start Manual Rev 2.45 - Software Versions 3.xx 8. Error Messages 8.1. Weighing Errors • Check Setup = This item can be checked on site by service personnel Error Description Resolution (U - - - - -) The weight is below the Increase the weight or minimum allowable weight decrease the minimum...
Quick Start Manual Rev 2.45 - Software Versions 3.xx 8.2. Setup and Calibration Errors Error Description Resolution The instrument may be in Access Full Setup to edit (ENTRY) Safe Setup and an item the item. (DENIED) that needs Full Setup has been selected for editing.
Page 13
Quick Start Manual Rev 2.45 - Software Versions 3.xx Error Description Resolution Scale wiring incorrect. (ZERO) An attempt has been (LO) made to calibrate zero below -2mV/V. An attempt has been Remove all weight from (ZERO) made to calibrate zero scale.
Quick Start Manual Rev 2.45 - Software Versions 3.xx 8.3. Diagnostic Errors • Check: This item can be checked on site by service personnel. • Return for Service: The instrument must be returned to the manufacturer for factory service. Error Description Resolution (E0001)