English 115 - Hach PHOSPHAX sigma Basic User Manual

Hide thumbs Also See for PHOSPHAX sigma:
Table of Contents

Advertisement

Available languages
  • EN

Available languages

  • ENGLISH, page 115
Measurement device technical data
Subject to changes.
Performance specifications
Reduction method, derived from DIN 38405 D11,
Measurement method
analysis photometric via IR LED photometer
Indoor or Outdoor use
Indoor only
0.01–0.50...5.00 mg/L P total phosphorus in 0.1 mg/L increments
Measurement ranges
0.01–0.50...5.00 mg/L P orthophosphate in 0.1 mg/L increments
(depending on design)
Measurement uncertainty
0.01 – 5.00 mg/L: ±2 % from the measurement value
Measurement interval
t
= 10 min
100
Sample requirement
approximately 100 mL/h
Ambient temperature
+5° C to +40° C, 95% relative humidity, non-condensing
Altitude
2000 m (6562 ft)
Calibration
Automatic at selectable intervals
Inspection interval
3 months
Maintenance requirement
approximately 1/2 hour per week, typical
Reagent stock
3 months (standard solution 6–12 months)
2 current outputs:
2 limit-value contacts:
Outputs
Service interface:
Interface:
Overvoltage category
II
Pollution degree
2
Mains connection
230 VAC ±10%, 50 Hz
Power consumption
Approximately 150 VA
Dimensions: W x H x D
560 mm x 810 mm x 330 mm
0/4...20 mA, maximum 500 Ω
potential-free, 24 V 1A
RS 232
bus-capable (optional)