Description

Specifications

Temperature Range
1
Temperature Stability
Cooling Capacity
2,3
4
Work Area Dimensions
(L x W x D)
Inches
Centimeters
Compressor
Heater
Working Volume
Gallons
Liters
Circulator Dimensions
(H x W x D)
Inches
Centimeters
Weight
Pounds
Kilograms
Pump
Pump Head
5
60 Hz
50 Hz
5
Flow
60 Hz
50 Hz

Section II General Information

The Thermo Scientific ULT 80 and ULT 95 are self-contained refrigerated
bath/circulators. The circulators consist of a circulating pump; heater; stain-
less steel reservoir; microprocessor temperature controller; and a cascade
configuration, air-cooled refrigeration system.
The refrigeration control is independent of the main power switch so that the
compressor may be turned off to allow fast warm-up. The automatic load
reset feature compensates the proportioning for changes in the bath load,
thereby eliminating shifts in setpoint accuracy.
ULT 80
-80°C to +10°C
± 0.03°C
250 watts at -70°C
5
3
/
x 7 x 9½
8
13.7 x 17.8 x 24.1
2 x 1 H.P.
1200 watts
47½ x 27
3
/
x 17¾
8
120.7 x 69.5 x 45.1
336
152.4
"Z" Pump (Force/Suction)
Max. Head 12' (3.7M)
Max. Head 10' 10" (3.3M)
0-10 liters/minute at 0' Head
0-10 liters/minute at 0' Head
1. 50 Hz ULT 95 circulators -85°C to -30°C. ULT 80 +80°C temperature range circulators are
available, these circulators have 2400 watt heaters (BOMs 180104201601 and 180106201601).
2. 200 watts for ULT 80, 50 Hz operation. 280 Watts for ULT 95, 50 Hz operation.
3. Cooling capacity is affected by ambient temperature - cooling capacity will decrease at the
rate of 1% per °F above 75°F.
4. Usable depth for ULT-80 is 8 inches.
5. Using fluid with specific gravity of 1.0.
- 5 -
ULT 95
-90°C to -30°C
± 0.2°C
340 watts at -80°C
2" DIA Fill Hole
5.1 cm Fill Hole
2 x 1.5 H.P.
1650 watts
4.0
15.1
48 x 32
1
/
x 21½
8
121.9 x 81.6 x 54.6
370
168.0
"H" Pump (Force only)
Max. Head 31' (9.4M)
Max. Head 21' (6.4M)
0-16.0 liters/minute at 0' Head
0-12.4 liters/minute at 0' Head