Omega Model CYD201/202 User’s Manual TABLE OF CONTENTS Chapter/Paragraph Title Page INTRODUCTION ................1-1 General ................1-1 Model CYD20X System Description.......1-1 INSTALLATION ................2-1 General ................2-1 Inspection and Unpacking..........2-1 Repackaging For Shipment ...........2-1 Front and Rear Panel Definitions ........2-2 2.3.1 Model CYD201 Definitions .........2-2 2.3.2...
Page 4
Omega Model CYD201/202 User’s Manual LIST OF ILLUSTRATIONS Figure No. Title Page Model CYD201, 202 Temperature Monitor Series ....1-3 Model CYD201 Front Panel ............2-2 Model CYD201 Rear Panel............2-3 Model CYD202S Front Panel............2-4 Model CYD202S Rear Panel ............2-5 Model CYD202D Front Panel............2-6 Model CYD202D Rear Panel ............2-7...
Omega Model CYD201/202 User’s Manual CHAPTER 1 INTRODUCTION GENERAL The CYD201, CYD202Series is designed and manufactured in the United States of America with these features: • Bright 4-Digit LED Display of Temperature in K or °C • Temperature Range: 1.4 K to 475 K (–271 °C to 202 °C) •...
Page 6
Omega Model CYD201/202 User’s Manual Diode sensors come in several packaging formats for easy mounting. Unlike a thermocouple, the lead wire in a diode measurement is not the active sensor component; using an additional connector to bring out the sensor signal from a cryogenic and/or vacuum system does not degrade the measurement.
Page 7
Omega Model CYD201/202 User’s Manual Model Model ° C ° C CYD201 CYD201 Omega Omega CYD201 Temperature Monitor CYD201 Temperature Monitor Model Model CYD202S CYD202S Sensor A Sensor A Sensor B Sensor B Omega Omega CYD202S Temperature Monitor CYD202S Temperature Monitor Channel A °...
Page 8
Omega Model CYD201/202 User’s Manual Table 1-1. Specifications THERMOMETRY Input/Display: Model CYD201 One input, one display Model CYD202S Two inputs, one display Model CYD202D Two inputs, two displays Measurement type: Two-lead Sensor Type: Silicon Diode Input Voltage Range: 5 Volts Sensor Excitation: 10 µA ±0.2%...
Inspect shipping containers for external damage. Make all claims for damage (apparent or concealed) or partial loss of shipment in writing to Omega Engineering within five (5) days from receipt of goods. If damage or loss is apparent, please notify the shipping agent immediately.
Omega Model CYD201/202 User’s Manual FRONT AND REAR PANEL DEFINITIONS This paragraph defines front and rear panels of the Model CYD201 (Paragraph 2.3.1), Model CYD202S (Paragraph 2.3.2), and Model CYD202D (Paragraph 2.3.3). 2.3.1 Model CYD201 Definitions The Model CYD201 has no front panel controls. The 4-digit red LED displays the temperature in Kelvin (K) or degrees Celsius (°C).
Omega Model CYD201/202 User’s Manual Model CYD201 Definitions (Continued) CAL Trimpot. The CAL trimpot adjusts instrument calibration. Refer to Paragraph 4.4 for details. Screw-Terminal Connector. The diode terminals are marked + for the anode and – for the cathode. VREF and Shield are used during unit calibration.
Omega Model CYD201/202 User’s Manual 2.3.2 Model CYD202S Definitions The Model CYD202S has one front panel control. The Sensor A/Sensor B switch selects which sensor data to display. Both sensors must conform to the same temperature response curve. The 4-digit red LED displays the selected temperature in Kelvin (K) or degrees Celsius (°C).
Omega Model CYD201/202 User’s Manual Model CYD202S Definitions (Continued) The Model CYD202S rear panel is the same as the Model CYD201 except: The upper screw-terminal connector applies to Sensor A. The lower screw- terminal connector applies to Sensor B. The front panel switch selects between the two sensors.
Omega Model CYD201/202 User’s Manual 2.3.3 Model CYD202D Definitions The Model CYD202D has no front panel controls. Two 4-digit red LEDs simultaneously display temperatures in Kelvin (K) or degrees Celsius (°C). In essence, the Model CYD202D operates as two independent units with a common power supply.
Omega Model CYD201/202 User’s Manual Model CYD202D Definitions (Continued) Descriptions of the Model CYD202D rear panel are the same as the Model CYD201 except two complete sets of DIP switches (refer to Paragraph 2.3.1), CAL trimpots, and screw-terminal connectors are provided. The upper set applies to Channel A.
SENSOR INSTALLATION RECOMMENDATIONS Abbreviated sensor installation recommendations for the Model CYD201 are included in this section. Please refer to the Omega Product Catalog for sensor specifications. Call Omega for copies of application notes or with questions or comments concerning sensor installation.
For temporary mounting in cold temperature applications, apply a thin layer ® of Apiezon N Grease between the sensor and sample to enhance thermal contact under slight pressure. The preferred method for mounting the sensor is the Omega CO Adapter. Installation...
Omega Model CYD201/202 User’s Manual Sensor Mounting (Continued) CAUTION: Omega will not warranty replace any device damaged by user- designed clamps or solder mounting. ® For semi-permanent mountings, use Stycast epoxy instead of Apiezon Grease. NOTE: Do not apply Stycast epoxy over the CY7-SD package--sensor stress may shift the readings.
To greatly reduce potential AC noise, connect twisted leads (pairs) between the measurement instruments and the diode sensors. Use 32 or 36 AWG Omega Engineering Duo-Twist™ Cryogenic Wire, which features phosphor bronze wire twisted at 3.15 twists per centimeter (8 twists per inch).
Curve 10. CY7 Series silicon diodes follow the same standard temperature response Curve 10 and are interchangeable. Curve 10 is programmed into all Omega Temperature Controllers, digital thermometers, and temperature monitors. Silicon Diode Series sensors come in five bands of tracking accuracy.
Omega Model CYD201/202 User’s Manual PANEL MOUNTING Panel mount cutout dimensions: Model CYD201 = 91 x 44 mm (3.58 x 1.72 in.); Model CYD202S & CYD202D = 91 mm square (3.58 in. square). Unplug the unit. Use 1/16 inch hex wrench to remove two headless screws.
Page 22
Omega Model CYD201/202 User’s Manual This Page Intentionally Left Blank 2-14 Installation...
Omega Model CYD201/202 User’s Manual CHAPTER 3 OPERATION GENERAL This chapter describes the front panel controls for each of the three models in Paragraphs 3.1 thru 3.3. TEMPERATURE UNIT SELECTION The temperature scale displayed on the front panel is selected by a DIP Switch labeled K/C on the rear panel.
Omega Model CYD201/202 User’s Manual FRONT PANEL OPERATION The Model CYD201 has no front panel controls. The 4-digit red LED displays the temperature in Kelvin (K) or degrees Celsius (°C). See Figure 2-1. The Model CYD202S has one front panel control. The Sensor A/Sensor B switch is used to select which of two sensors temperature data is displayed.
Omega Model CYD201/202 User’s Manual CHAPTER 4 SERVICE GENERAL This chapter cover General Troubleshooting (Paragraph 4.1), Operating Software Replacement (Paragraph 4.2), the Model CYD201 Printed Circuit Board (PCB) (Paragraph 4.3), and Calibration (Paragraph 4.4). GENERAL TROUBLESHOOTING If there is no power, first verify that 24 VAC is present. Next, check the fuse.
Omega Model CYD201/202 User’s Manual PCB DESCRIPTION Below is a brief description of major components on the Model CYD201/2 Printed Circuit Board (PCB). See Figure 4-1. Microprocessor/ROM (U10). This integrated circuit is a combination microprocessor, program memory, and Precision Option Curve (if present).
Page 28
Omega Model CYD201/202 User’s Manual This Page Intentionally Left Blank Service...