Celectric/Gas Connections And Installation Of The Work Surface - D FASTER CytoFAST Top Operating And Maintenance Manual

Vertical laminar airflow cabinet for cytotoxic drugs
Table of Contents

Advertisement

a D:GROUP Company
2.C
ELECTRIC/GAS CONNECTIONS and INSTALLATION OF THE WORK SURFACE
The electrical connection of the CytoFAST Top Biohazard cabinet is made by connecting the power cable
located on the upper of the right side of the cabinet to a suitable power point (see technical table) . When the
cabinet is connected, the green light on the control panel switches on (see chapter 4D).
If stipulated by local legislation, insert upstream of the power line an automatic protection overload switch
provided with a differential relay, with a rated switching voltage no greater than 30 mA.
The cabinet can be provided with two gas/vacuum intakes (optional). The connection with the two
gas/vacuum intakes is made according to the type of connection: town gas or industrial gas (air, vacuum,
nitrogen, etc.).
The cabinet must be connected with the town gas mains through an approved conduit for safety reasons.
For the installation of the work surface, proceed as follows:
-
remove the protective paper from the work surface resting against the back of the cabinet, taking care
not to scratch its surface,
-
remove the padding/sterilization plate,
-
open the shatterproof front window,
-
clean the work surface with a damp cloth soaked in alcohol or soapy water or with a commonly available
product designed for stainless steel,
-
place the table in the work chamber, allowing the back to slide on the chamber's supporting bases up to
its bottom wall,
-
close the shatterproof front window.