General - D FASTER CytoFAST Top Operating And Maintenance Manual

Vertical laminar airflow cabinet for cytotoxic drugs
Table of Contents

Advertisement

a D:GROUP Company
1

GENERAL

Vertical laminar airflow benches with partial exhaust and protection barrier, the CytoFAST Top cabinets, are
designed to protect both the material to be manipulated from contamination and to protect the operator and
the environment from microbial cytotoxic contamination hazards.
The de-dusted, filtered and sterile air passing through the main HEPA filter ensures optimum airflow
laminarity on the work surface, thanks to its even perforation and the frontal air barrier. The system is
designed to create a protective barrier and prevent exchanges between the internal, potentially biologically
contaminated air, and the outside.
CytoFAST Top cabinets are Class II Biohazard cabinets, according to the definitions now adopted by all the
main international standards, with about 70% laminar flow recirculated air and 30% extracted air with prior
HEPA filtering. Suitable for the treatment of class CDC-1/2/3, DHSS-C/B1/B2 group 1,2,3 microorganisms as
per D.L. 81/08 "Safety in the working environment" and wherever a product-personnel-environment cross-
protection is required.
Especially suitable for applications such as:
-
Manipulation of eziologic agents having a known pathogenicity on humans and animals.
-
Presence of high concentrations of biologic materials.
-
Presence of agents likely to cause genetic alterations or synergetic activities with other materials.
-
Oncogenic viruses.
But above all designed for:
-
Preparation and handling of cytotoxic drugs
-
Preparation and manipulation of antineoplastic chemotherapeutics
Do not manipulate the products on the slotted part at the entry of the chamber, but in the middle of the work
surface.
The performances of the cabinets are detailed in the TESTING CERTIFICATE below, in compliance with the
requirements of:
-
ISO 14644-1 Class 5
For the Biohazard/Cytotoxicity:
-
DIN 12980: 2005
-
EN 12469: 2000
Furthermore CytoFAST Top cabinets fulfil the armonized standards EN 61010-1 as well as EN 61326
according to the applicable European directives regarding the CE marking.
CytoFAST Top cabinets comply with the above-mentioned standards ONLY if the instruments connected to
the electrical socket positioned inside the work chamber are "CE" marked or in any case it meets the above
mentioned standards aiming to avoid any electromagnetic interference.
All FASTER's cabinets are provided with high insertion loss filters.
Faster s.r.l. cannot be held responsible for malfunctions, damage to people or property due to non-
compliance, poor or no maintenance, or improper use of the cabinet.