D FASTER CytoFAST Top Operating And Maintenance Manual page 5

Vertical laminar airflow cabinet for cytotoxic drugs
Table of Contents

Advertisement

a D:GROUP Company
Thimble method example
LEGENDA:
1. Exhaust air from the cabinet.
2. Air from the environment.
÷
3. Bleed Air (100
200 m
flow rate from the cabinet) to a dedicated exhaust
fan (in the electronic board a voltage free contact is
available –see electrical diagram- to check when
ventilation is ON).
For the connection to the gas and/or vacuum networks, read carefully chapter 2C.
The quantity of heat generated by the cabinet, if the air is not extracted ouside the room, is the following for
the eight models respectively:
The quantity of heat generated by the cabinet, if the air is not extracted outside the room, is the following for
the four models respectively:
-
CytoFAST Top 209:
-
CytoFAST Top 212:
-
CytoFAST Top 215:
-
CytoFAST Top 218:
The installation must be done by technicians authorized by Faster S.r.l. or by the distributor
00055EN Rev.10 – 12/2011
3
/h more than the exhaust air
410 Kcal/h
460 Kcal/h
520 Kcal/h
540 Kcal/h
1
Microbiological Safety Cabinet
3
THIMBLE
2