1.0 Introduction
Congratulations on purchasing your Labnet Pro 30 Laboratory Reciprocal Shaker. This shaker features
a left to right linear action to provide gentle to vigorous shaking. Four platform configurations make this
model valuable for a variety of applications.
2.0 Packing
The Labnet Pro 30 Laboratory Reciprocal Shaker includes the following components:
Pro 30 Reciprocal Shaker
Instruction Manual
NOTE: Platform needs to be purchased separately.
3 .0 Specifications
Electric Power Supply
Fuse
Speed
Stroke Length
Timer
Load
Dimensions (W x D x H)
Platform Dimensions (W x D)
Weight
Operating Conditions
4.0 Safety Information
Do not use the shaker near water sources. Be sure that water doesn't get on the device, especially
during the cleaning procedures.
Do not expose the shaker to aggressive vapors (e.g. organic solvents).
Do not insert objects into the shaker (such as screw driver or metal rods).
Never operate the shaker without platform securely attached.
Hazardous samples should be shaken only in appropriate containment vessels.
NOTE: If the equipment is used in a manner not specified by the manufacturer, the protection provided
by the equipment may be impaired.
Do not shake flammable or explosive samples.
115V ± 10%, 50/60Hz, 0.2A or 230V ± 10%, 50/60Hz, 0.1A
230V includes EU and UK power cords
2 x 0.5A (115V); 2 x 0.25A (230V)
From 5 to 150 strokes per minute
30 mm
20 to 120 minutes, in 10 minute steps, HOLD function
11 lbs. (5 kg) including platform
10 x 12.6 x 6.3 in. (25.5 x 32 x 16 cm)
12 x 12 in. (30 x 30 cm)
15 lbs. (6.6 kg)
4°C to 70°C, up to 85% RH, non-condensing
Instruction Manual
|
1
Need help?
Do you have a question about the Pro 30 and is the answer not in the manual?