Philips 9890 010 83921 Service Manual page 20

Table of Contents

Advertisement

BuckyDiagnost FS IsoRAD
Note
Do not touch the fixing screws (1).
S Solder the new alpha switch plate at the wires
according the following table.
-- Pay attention to the diode (2).
Labeled wires connected to:
UEX4:02
1
UEX4:08
3
UEX4:09
4
UEX4:04
2
S Insert and fix the alpha switch plate.
S Hook the spring.
S Switch the system ON.
S Ensure the 0°, +90° and --90° positions at control
handle according to table.
-- The LED light is on only in 0°, +90° and --90°
positions.
S1
S2
ON
ON
ON
OFF
OFF
ON
S Re--install the cover.
1.3.6. Replacement of the transverse brake
Code no. 4512- -130- -7291.
S Switch the system OFF.
S Remove the bottom cover of the tube carrier arm.
4--8
to the cathode of both
diodes
to UFA S1:1
to UFA S1:4
to UFA S2:1
degree
0
+90
--90
3
4
1 h
(04.0)
E
2004 Philips Medical Systems
ALL RIGHTS RESERVED
REPLACEMENTS
1
2
1
2
Replacement of the
transverse brake
BuckyDiagnost FS IsoRAD
4_Replacements_mit_12NC_04

Advertisement

Table of Contents
loading

Table of Contents