REPLACEMENTS
S Pull out the tube carrier arm to its maximum size.
Warning!
If the end stopper is removed the tube
carrier arm can fall down. Secure the
tube carrier arm, e.g. by lifting the table.
S Remove the end stopper by loosening the screws
(1).
S Remove the screws (1) from the brake plate.
Note
Keep in mind the positions of the cable!
S Remove the cable (2).
S Replace the brake (3) at the brake plate.
S Solder the cable.
S Fix the cable at the tube carrier arm.
S Fix the end stopper.
S Check the function of the brake.
S Re-install the cover.
BuckyDiagnost FS IsoRAD
4_Replacements_mit_12NC_04
1
2
1
(04.0)
E
2004 Philips Medical Systems
ALL RIGHTS RESERVED
BuckyDiagnost FS IsoRAD
3
4--9