Philips 9890 010 83921 Service Manual page 27

Table of Contents

Advertisement

BuckyDiagnost FS IsoRAD
1.4.2. SID laser alignment
S Select the RGDV4 for free exposure (1), see
SID-Selection with Program X-Scope, function
None should not be selected!
S Ask the operator for the prefered SID. The default
pre--adjustment is SID = 1100 mm.
S Raise the table top into the prefered value of the
SID = 900 ... 1200 mm.
S Press button (2) = center laser and pulsing SID
laser and diaphragm light are ON.
S Observe the center laser and the pulses SID laser
on the table top.
-- If the beams are not merged, loosen the two
fixing screws (3) and adjust the SID laser with
the eccenter screw (4).
-- Tighten the two fixing screws (3).
1.4.3. Final work
S Install the removed control handle grip on the rear
side
Note
Pay attention to the laser wires, that they are not
damaged by attaching the grip! If necessary, fix the
wires with adhesive tape.
BuckyDiagnost FS IsoRAD
6_Adjustments
1
2
Rear side view
(04.0)
E
2004 Philips Medical Systems
ALL RIGHTS RESERVED
ADJUSTMENTS
4
3
3
6--3