BuckyDiagnost FS IsoRAD
1.3.
Long tube carrier arm, extendable
1.3.1. Replacement of the complete arm
Code no. 4512- -131- -8753.
1.3.2. Replacement of the base plate installed micro switch
Code no. 4512- -133- -0544.
S Remove the cover at the bottom of the tube carrier
arm.
S Remove the connector (1).
S Loosen the four screws (1) and replace the base
plate installed micro switch.
-- Ensure the correct function of the micro switch.
S Check the function of the system.
S Re-install the cover.
4--4
4 h
Replacement of the base plate
0.5 h
1
(04.0)
E
2004 Philips Medical Systems
ALL RIGHTS RESERVED
REPLACEMENTS
Replacement of the
complete tube carrier arm
installed micro switch
1
BuckyDiagnost FS IsoRAD
4_Replacements_mit_12NC_04