BuckyDiagnost FS IsoRAD
1.3.4. Replacement of the magnetic unit
Code no. 4512- -133- -0584.
S Remove the cover at the bottom of the tube carrier
arm.
S Remove the base plate installed micro switch
according to chapter 1.3.2.
S Loosen the four screws (1) and replace the
magnetic unit.
S Install the base plate installed micro switch
according to chapter 1.3.2.
1.3.5. Replacement of the alpha switch plate
Code No. 4512 13 1441.
S Switch the system OFF.
S Turn the control handle clockwise to its end
position.
S Remove the cover (1).