Table 1 (support for IEC 60601-1-2:2007 and IEC 60601-1-2:2014)
Guidelines and manufacture's declaration - electromagnetic emissions
This device is intended for use in the electromagnetic environment specified below.
The customer or the user of this device should assure that it is used in such an environment.
Emissions test
Compliance
RF emissions
Group 1
CISPR 11
RF emissions
Class B
CISPR 11
Harmonic emissions
Class A
IEC 61000-3-2
Voltage fluctuations/flicker
emissions
Complies
IEC 61000-3-3
IMPORTANT
• If I/F Cable4 1.5mD, I/F Cable4 8mD, and/or I/F Cable4 8mU is used, the emissions characteristics of this device is CISPR 11
class A that is suitable for the industrial areas and the hospitals. If it is used in a residential environment, for which CISPR 11
class B is normally required, this equipment might not offer adequate protection to radio-frequency communication services, and
it may be necessary to take a mitigation measures, for example, by relocating or re-orientating the device.
Table 2 (support for IEC 60601-1-2:2007)
Guidelines and manufacturer's declaration - electromagnetic immunity
This device is intended for use in the electromagnetic environment specified below.
The customer or the user of this device should assure that it is used in such an environment.
Immunity test
IEC 60601 test level
±6 kV contact
Electrostatic discharge
(ESD)
IEC 61000-4-2
±8 kV air
±2 kV for power supply
Electrical fast transient/
lines
burst
IEC 61000-4-4
±1 kV for input/output lines ±1 kV for input/output lines
±1 kV line(s) to line(s)
Surge
IEC 61000-4-5
±2 kV line(s) to earth
<5 % U
for 0.5 cycle
Voltage dips, short
40 % U
interruptions and voltage
for 5 cycles
variations on power sup-
70 % U
ply input lines
for 25 cycles
IEC 61000-4-11
<5 % U
for 5 sec
Power frequency
(50/60 Hz) magnetic field
3 A/m
IEC 61000-4-8
[NOTE] U
is the AC mains voltage prior to application of the test level.
T
Electromagnetic environment - guidelines
The device uses RF energy only for its internal function. Therefore, its RF emissions
are very low and are not likely to cause any interference in nearby electronic equip-
ment
This device is suitable for use in all establishments including the following:
Domestic establishments and those directly connected to the public low-voltage
power supply network that supplies buildings for domestic purposes.
Compliance level
±6 kV contact
±8 kV air
±2 kV for power supply
lines
±1 kV line(s) to line(s)
±2 kV line(s) to earth
(>95 % dip in U
)
<5 % U
(>95 % dip in U
T
T
T
for 0.5 cycle
(60 % dip in U
)
40 % U
(60 % dip in U
T
T
T
for 5 cycles
(30 % dip in U
)
70 % U
(30 % dip in U
T
T
T
for 25 cycles
(<95 % dip in U
)
<5 % U
(<95 % dip in U
T
T
T
for 5 sec
3 A/m
21
Safety precautions
Electromagnetic environment - guide-
lines
Floors should be wood, concrete or ceramic
tile. If floors are covered with synthetic
material, the relative humidity should be
at least 30 %. Mains power quality should
be that of a typical commercial or hospital
environment.
Mains power quality should be that of a typi-
cal commercial or hospital environment.
)
T
Mains power quality should be that of a
typical commercial or hospital environment.
)
T
If the user of the device requires continued
operation during power mains interruptions,
)
T
it is recommended that the device be pow-
ered from an uninterrupted power supply or
a battery
)
T
Power frequency magnetic fields should be
at levels characteristic of a typical location
in a typical commercial or hospital environ-
ment