General
Fig. 2:
V
With housing covers open and safety covers removed.
Back:
Left side:
Right side:
Siemens AG
Medical Solutions
Locations with dangerous voltage
CAUTION: DC voltage (300 V) from the battery block to PCB D982!
Always disconnect the battery plug from BK1-BK4.
CAUTION: DC voltage (440V) at capacitor bank (D927)
CAUTION: DC voltage (440V) directly at the capacitor bank!
CAUTION: DC voltage (440V) directly at the capacitor bank!
CAUTION: AC voltage (> 100 V or > 60 V - half line voltage) at power sup-
plies U1 and U2 when the power cord is connected.
SPR8-230.814.30.03.02
11.05
MOBILETT XP Digital
Page 11 of 70
CS SD 24
11