Information On Measuring The Leakage Current - Siemens MOBILETT XP Digital Installation And Startup

Hide thumbs Also See for MOBILETT XP Digital:
Table of Contents

Advertisement

22

Information on measuring the leakage current

Observe the instructions in the "Safety Rules for Installation and Repair"
(ARTD-002.731.17 ...).
WARNING
Initial measurement
Perform the leakage current measurement after completion of all work.
Perform the measurement according to DIN VDE 0751, Part 1 (see ARTD-002.731.17....),
and record the determined value as the first measured value.
Measurement of the leakage current according to the differential current method (mea-
surement setup according to
is not dangerous to the person performing the measurement and other persons.
However, please note the minimum resolution of the leakage current measuring device
and any additional manufacturer information restricting the use of the measuring device.
Fig. 7:
Pos. 1
Pos. 2
Pos. 3
MOBILETT XP Digital
Electrical voltage!
Non-compliance can lead to severe injury and even death.
The leakage current measurement may be performed on
systems of protection class I only after the protective
conductor test has been passed.
(Fig. 7 / p.
Measuring circuit for measuring the system leakage current according to the differential
current method in compliance with DIN VDE 0751-1/2001-10, Fig. C6 for protection
class I.
= System
= Application part type B (if available)
= Measurement setup (integrated into measuring device)
SPR8-230.814.30.03.02
11.05
General start-up information
22)) must be given preference, since this method
Page 22 of 70