Repair Procedure 1.2. Power ON 1. Examining the Internal Components First 2. Inspecting The value, S.C.A. Prism Normal State Checking Abnormal State Cleaning the Pinhole Power off and on Normal State Checking Abnormal State Setup Again Exit Figure 4. Repair procedure flow chart...
Auto Lensmeter HLM-7000P Cautions 1.3. Always protect this instrument with Dust Cover after using. Do not vibrate or drop this instrument; it can cause damage. Use soft cloth or cotton swaps with alcohol when clean the pinhole.
Collimation lens : makes parallel light. Signal Receiving Part Hartmann plate : divide the signal into 81 points. Prevent the other light except the signal 2-dimentional CMOS sensor : converts signals into digital value. Measurement Principle 1.6. Basic Principle When there is no lens (0D state), parallel light that has transmitted the collimator lens passes the pinhole and makes image on CCD.
Auto Lensmeter HLM-7000P 2. Checking and Setup Method Setup Order 2.1. Start up with pressing 3 and 6 keys, then you can meet “Setup Mode” as below. While this mode, ‘SETUP MODE’ mark will show in the upper area. Calibration can be processed only in SETUP mode; LENS, PRISM, CYLINDER, PD, UV.
How to Set Origin 2.2. The steps of setting origin are as follows: A. Press the ‘LENS’ (Key 1) button. B. Press the ‘MEASURE’ (Key 2) button. C. Determine the lens to measure by using two ‘-INDEX’, ’+INDEX’ buttons. D. ‘D-0.01’ button and ‘D+0.01’ button by the given value in the Standard Lens-set specification.
Auto Lensmeter HLM-7000P How to Set 12 standard lenses 2.3. If the measured value is not accurate, you should set the diopter variables again by using the standard lens-set: for all the 12 lenses in the standard lens set. Always keep the standard lens clean out of dust or stain.
How to Set Prism 2.4. If the measured value is not accurate, you should set the 2, 5, 10, 15, 20 lens again by using the standard lens-set. A. First, convert prism display format into P-B. B. Then, press the ‘PRISM’ (Key 2) button. C.
Auto Lensmeter HLM-7000P ANGLE PRISM BASE 0° ~ 30° center 31° ~ 60° center 61° ~ 90° center 91° ~ 120° center 121° ~ 150° center 151° ~ 180° center 181° ~ 210° center 211° ~ 240° center 241° ~ 270° center 271°...
Page 15
C. ‘-0.01’ button and ‘+0.01’ button by the given value in the Standard Lens-set specification. D. Locate it to the center while cling it to the lens. (Axis : 90°, Prism_x, y: 0) E. Press the ‘MEASURE’ button and check if ‘A90’ value is changed. F.
Auto Lensmeter HLM-7000P PD Setup 2.6. With the PD Sensor, we can calibrate the PD value of our machine. The procedures are as followings: A. Press the ‘PD’ ( ) button. B. Push in the PD sensor is the leftmost. And press the ‘L Lim’ button.
UV Setup 2.7. With the UV, we can calibrate the UV value of our machine. The procedures are as followings: A. Press the ‘UV’ ( ) button. B. Open UV cover. C. Focus to UV 0.0 by using ‘INDEX’ button. D.
Auto Lensmeter HLM-7000P Diopter Setting Variables List 2.8. The followings are diopter setting variables list: WAVEFRONT DATA INFORMATION [ALL DIOPTER] 0 [-25.01]: L=+174.05; 123.48, 122.67: average distance and distance rates between the four points and -25D 1 [-19.99]: L=+168.52; 119.09, 119.23: average distance and distance rates between the four points and -20D …...
How to change the CCD camera 2.9. The procedures for changing a camera are as followings. A. Remove all subassemblies that are the side cover, the lens table and the lower cover and the CCD cover etc. B. Loose the Base Screw I and then take out the old CCD assembly. C.
Page 20
Auto Lensmeter HLM-7000P We should place the CMOS Camera keeping two conditions: The first condition is the position of center point. The center position is shown at the 5 row like this. - CEN: (+640.00, +512.00) It must be within ( +640.00 ± 15, 512 ± 15 ) The second one is the orientation of CMOS Camera.
2.10. How to change the LED Assembly The spare part for the LED assembly is the Upper Frame. So, we provide all upper frame assembly for the LED assembly. The procedures for changing the upper frame Assembly are as followings: A.
Auto Lensmeter HLM-7000P 2.11. How to change the Pinhole Housing The procedures for changing the Pinhole Housing are as followings: 1. Take out the Lens Supporter. 2. Remove the Pinhole Housing Cover which is the below cover under the Lens Supporter.
Page 26
Auto Lensmeter HLM-7000P Component Removal method BASE FRAME - Disconnect all cables from [14]. FRAME CR RUBBER 1 - Remove 4 screws and separate [14]. FRAME CR RUBBER 2 - Remove 4 screws and separate [8]. - Remove 4 screws, separate [5] and extract [7].
Page 28
Auto Lensmeter HLM-7000P Component Removal method BACK BONE - Disconnect ‘PD Bar Board’ cable from ‘Main Board’. B6_LED-BONE-ASSY - Remove cap [18] SHAFT FIX BRACKET - Reference : 3.3.1 PD-Bar Sub-Assembly Please see the drawing for disassemble, as needed GUIDE SHAFT 4...
Page 30
Auto Lensmeter HLM-7000P Component Removal method PD BAR PLATE - First, remove [6] from Back Bone Assembly and separate B4_PD-SHAFT [7] [Washer]. PD RACK GEAR SHAFT - Next, remove 2 screws and separate [20] 접시머리 나사 (3*10) - Next, remove 4 screws and take off [17]...
Page 32
Auto Lensmeter HLM-7000P Component Removal method FOOT HOLDER Please see the drawing for disassemble, as needed F T2 CONTROL POS PIN POSITION FOOT SPRING - Assembly is the reverse procedure of disassembly. C7_POSITION-FOOT-RUBBER- ASSY B4_FOOT-FRAME-BRACKET-13...
Page 34
Auto Lensmeter HLM-7000P Component Removal method B4_PEN-FRAME-BRK Please see the drawing for disassemble, as needed B4_PEN-PRESS-BRK B4_PEN-CONT-SUPPT - Assembly is the reverse procedure of disassembly. 접시머리 나사 (2.6*6) 무두볼트(SS) (3*12) 흑착색(3가 크롬도 금) WRENCH BOLT PEN GUIDE SHAFT PEN GUIDE SPRING E-RING 육각...
Page 36
Auto Lensmeter HLM-7000P Component Removal method B4_PEN-SLIDE-BASE Please see the drawing for disassemble, as needed B4_SLIDE-ASSY - Be careful the balls PCB SUPPORT BOLT - Assembly is the reverse procedure of disassembly. 접시머리 나사 (3*5) 무두볼트(SS) (3*4) WRENCH BOLT...
Page 40
Auto Lensmeter HLM-7000P Component Removal method B4_LED-BONE-13 -Turn part [2] for disassemble lens [3] B4_LED-LENS-NUT-13 Please see the drawing for disassemble, as needed NEW_LM_LENS B6_LED-HOUSING-ASSY-GREEN - Assembly is the reverse procedure of disassembly 무두볼트(SS) (3*5)
Page 44
Auto Lensmeter HLM-7000P Component Removal method UV BONE - Disconnect cables from [5] [8]. UV TEFRON(L) - Remove 2 screws from [5] and separate [5] [3]. UV TEFRON(U) - Remove 4 screws from [8] and separate [8] [2]. HARNESS(MB TO UV&BLUE LED)
Page 46
Auto Lensmeter HLM-7000P Component Removal method COVER TOP 5 - Disconnect cables from [4]. COVER MEM BUTTON PLATE - Remove 4 screws from [1] and separate [2] [3]. COVER MEM BUTTON PCB ASSY(HLMM-100) - Be careful that there is no scratch and assembly is the MEM LED DIFFUSION SHEET reverse procedure of disassembly.
Inspecting SMPS 4.2. Check1. Check if POWER SWITCH and RUG is connected properly. Check2. Check FUSE in POWER SOCKET with DMM. Check1. Is Power LED on in SMPS MODULE? Check2. Check the voltage of DC POWER Cable with DMM. Change SMPS MODULE (It should be AC100 - 220V) *REFERENCE: “ELECTRICAL CONSTRUCTION DIAGRAM”...
Auto Lensmeter HLM-7000P LCD TEST 4.3. Measurement screen is not displayed… Check1. SMPS TEST Check1. LCD is bright but there is no Check if connector of LCD is character or figure. connected correctly to MAIN Check2. There are horizontal lines on LCD BOARD.
UV MODULE TEST 4.5. Check UV-LED states Change UV-LED Check1. Check if light radiated from OPTIC PCB ASS’Y LED comes out from UV measurement part. UV-LED values in UV mode should change(100% -> 0%). Check1. Check if UV SENSER in UV measurement part is hidden by obstacles such as FILTER COVER.
Auto Lensmeter HLM-7000P PD TEST 4.6. PD value is not displayed… Check1. Check right/left switching when you move Change MAIN LEVER right and left compared to the center of PD. BOARD Change PD MODULE...
THERMAL PRINTER TEST 4.7. “PRT: HEAD UP” message is displayed on LCD… Check1. Open PRINTER COVER and check if HEADER LEVER towards upside. “PRT: PP EMPTY” message is displayed on LCD when pressing PRINTER BUTTON… Check1. Open PRINTER COVER. Check2.
5. How to upgrade the firmware software 5.1. Introduction Upgrade software for HLM-7000P proceeds by serial communication. DNW which is PC program to transmit firmware software from PC to Lensmeter is necessary. An administrator of the Lensmeter hardware can receive the Firmware by e-mail or else and load it to his/her lensmeter by this downloading program when the new upgrade version of firmware comes out.
5.2. How to upgrade the firmware software using DNW application in HLM machines A. Execute the DNW application. B. Press ‘Options’ button to set the COM port.
Page 56
Auto Lensmeter HLM-7000P C. Set the connected COM port and baud rate (115200). D. Press ‘Connect’ button.
Page 57
E. Power on the Lensmeter with pressing 5 key. Then you can see the below messages. F. Press ‘Transmit’ button and select firmware file.
Page 58
Auto Lensmeter HLM-7000P G. . Wait until download is the end. H. Reboot the Lensmeter after restart system message.