Installation - Thermo Scientific 88881-003 User Manual

Bottle/tube roller

Advertisement

Section 1
Unpacking and Installation

Installation

1-2
Bottle/Tube Roller
1. Control box installation
This unit is equipped with a separate control box. Fasten the nut on the
connection plug of control box to the unit.
Note Keys on the panel of the unit will become invalid when control box
is connected. One control box controls one unit only.
2. Roller bars removal and installation
Roller bars can be removed and installed for different sized tubes.
2.1 Roller bars removal
Counterclockwise twist the knob to be loose on the unit end(see
Figure 1-1) and then uplift the pressure sheet-metal. Uplift the
roller bar (see Figure 1-2) and remove it. Press the sheet metal to
the original position and clockwise twist the knob for fastening.
Figure 1-1. Loosen the Knob
2.2 Roller bars installation
Counterclockwise twist the knob to be loose on the unit end (see
Figure 1-1) and then uplift the pressure sheet-metal. Aim hexagon
hole of roller at hexagon shaft of the unit and press the other head
into semicircle orifice. Press the sheet metal to the knob position
and clockwise twist the knob for fastening.
3. O-Rings
When rolling test tubes, tubes, tubes may roll off the
bars due to the larger outer diameter of the tube cap.
Place o-rings on the other end of the tube (see
Figure1-3) to prevent the tube from falling off. Two
o-rings for each tube are recommended.
Figure 1-2. Lift Roller Bar
Figure 1-3. O-rings