Trouble-Shooting; Errors And Warnings - Endress+Hauser Ceraphant T PTC31 Operating Instructions Manual

Process pressure measurement
Hide thumbs Also See for Ceraphant T PTC31:
Table of Contents

Advertisement

Ceraphant T PTC31, PTP31, PTP35
9

Trouble-shooting

9.1

Errors and warnings

If an error in the device occurs, the colour of the status LED changes from green to red and the
background illumination of the digital display changes from white to red. The display shows:
• E-code for errors. In the event of an error message, the measured value is uncertain.
• W-code for warnings. In the event of a warning, the measured value is reliable.
E-Code
E011
E012
E019
E015
E020
E021
E022
E025
E026
E040
E042
E044
W-Code
W107
W202
W209
W210
W212
Endress+Hauser
Explanation
Device configuration defective
Error in measurement or
pressure out of specification
Power supply outside specification
Memory error
Power is only supplied to the device via the
communication interface (measurement is
deactivated)
Switching contact 1 is not open although it
should be
Switching contact 2 is not open although it
should be
VCC (controller voltage) is out of working area
Output current can no longer be generated (only
for 4 to 20 mA output, e.g. load at analog output
too high or open analog output)
Output current drifts too much (±0.5 mA)
Explanation
Simulation active
Pressure outside the sensor range
Device starts
Configuration modified (warning code will be
displayed for 15 s approx.)
Sensor signal outside the permitted range
Remedy
Reset device (see chapter 5.1.5)
Check pressure, return device to E+H where
necessary
Check operating voltage
Return device to Endress+Hauser
Check operating voltage
Switching contact defective.
Return device to Endress+Hauser
Switching contact defective.
Return device to Endress+Hauser
Return device to Endress+Hauser
Check load; switch off analog output via
configuration if it isn't required (see chapter
5.1.4)
Return device to Endress+Hauser
Remedy
Switch off the output simulation for output 1 and
output 2
Operate the device in the specified pressure range
Operate the device in the specified pressure range
Trouble-shooting
27

Advertisement

Table of Contents
loading

This manual is also suitable for:

Ceraphant t ptp31Ceraphant t ptp35

Table of Contents