Endress+Hauser Ceraphant T PTC31 Operating Instructions Manual

Endress+Hauser Ceraphant T PTC31 Operating Instructions Manual

Process pressure measurement
Hide thumbs Also See for Ceraphant T PTC31:

Advertisement

Quick Links

Operating Instructions
Ceraphant T PTC31, PTP31, PTP35
Process pressure measurement
KA225P/00/EN/10.09
71102965
Release 1.04

Advertisement

Table of Contents
loading

Summary of Contents for Endress+Hauser Ceraphant T PTC31

  • Page 1 Operating Instructions Ceraphant T PTC31, PTP31, PTP35 Process pressure measurement KA225P/00/EN/10.09 71102965 Release 1.04...
  • Page 2: Table Of Contents

    Table of contents Ceraphant T PTC31, PTP31, PTP35 Table of contents 1 Safety instructions ..........3 1.1 Designated use .
  • Page 3: Safety Instructions

    Ceraphant T PTC31, PTP31, PTP35 Safety instructions Safety instructions Designated use The Ceraphant T is a pressure switch for measuring and monitoring absolute and gauge pressures. The device has been safely built with state-of-the-art technology and meets the applicable requirements and EC Directives. It can, however, be a source of danger if used incorrectly or for anything other than the designated use.
  • Page 4: Identification

    Identification Ceraphant T PTC31, PTP31, PTP35 Identification Nameplate To identify your device, compare the complete order code and the version information on the delivery papers with the data on the nameplate. Order Code: Mat: Ser.-No.: I out: TAG: Rel.: MWP: L–...
  • Page 5: Ce Mark, Declaration Of Conformity

    The device complies with the applicable standards and regulations as listed in the EC declaration of conformity and thus complies with the statutory requirements of the EC Directives. Endress+Hauser confirms the successful testing of the device by affixing to it the CE mark.
  • Page 6: Installation

    Installation Ceraphant T PTC31, PTP31, PTP35 Installation Incoming acceptance, storage • Incoming acceptance: Check the packaging and the device for damage. Check that the goods delivered are complete and nothing is missing. • Storage: Storage temperature -40...+85 °C (-40 to +185 °F)
  • Page 7: Post-Installation Check

    Ceraphant T PTC31, PTP31, PTP35 Installation 3.2.1 Mounting instructions • Do not mount the device in the product flow or at a point where it could be affected by pressure pulses • Calibration and functional testing are easier if the device is mounted downstream of a shut-off assembly •...
  • Page 8: Wiring

    Wiring Ceraphant T PTC31, PTP31, PTP35 Wiring DC voltage version with M12 connector A2’ ➀ L– L– A3’ 4...20mA Diag. 4...20mA L– L– P01-PTx3xxxx-04-xx-xx-xx-002 Fig. 3: Ceraphant T with M12x1 connector A1: 1x PNP switch output A2: 2x PNP switch outputs R1 and m (R2) A2’: 2x PNP switch outputs R1 and m (diagnosis/break contact with adjustment "DESINA")
  • Page 9: Dc Voltage Version With Cable

    Ceraphant T PTC31, PTP31, PTP35 Wiring DC voltage version with cable C2’ ➀ L– L– GNYE GNYE C3’ 4...20mA 4...20mA Diag. L– L– GNYE GNYE P01-PTx3xxxx-04-xx-xx-xx-004 Fig. 5: Ceraphant T with cable connection C1: 1x PNP switch output C2: PNP switch outputs R1 and m(R2) C2’: PNP switch outputs R1 and m (diagnosis/break contact with adjustment "DESINA")
  • Page 10: Operation

    Operation Ceraphant T PTC31, PTP31, PTP35 Operation On-site operation The Ceraphant T is operated by means of three keys. The digital display and the light emitting diodes (LED) support navigation in the operating menu. Operating key Digital display LED for status Green = ok 22.07...
  • Page 11 Ceraphant T PTC31, PTP31, PTP35 Operation 5.1.1 Navigating in the operating menu b a r > 3 s SAVE > 3 s P01-PTx3xxxx-19-xx-xx-xx-005 Fig. 7: Navigating in the operating menu A Function group selection B Function selection ➀ Enter the operating menu –...
  • Page 12 Operation Ceraphant T PTC31, PTP31, PTP35 5.1.2 Structure of the operating menu The structure below shows all the possible fields of the operating menu. P01-PTx3xxxx-19-xx-xx-xx-100 Fig. 8: Operating menu: A function groups, B functions, C settings Endress+Hauser...
  • Page 13: Operation With Personal Computer

    Ceraphant T PTC31, PTP31, PTP35 Operation Operation with personal computer The device can be configured with the configuration software ReadWin® 2000 or FieldCare. For the connection between the USB port of the computer and the device a configuration kit (e.g. TXU10 or FXA291)) is necessary.
  • Page 14 Operation Ceraphant T PTC31, PTP31, PTP35 5.2.1 Additional operating options In addition to the operating options listed in the previous "On-site operation" section, the ReadWin 2000 or FieldCare configuration software provides further information on the Ceraphant T: Function group Description...
  • Page 15: Commissioning

    Ceraphant T PTC31, PTP31, PTP35 Commissioning Commissioning Function check Carry out a post-installation and a post-connection check as per the checklist before commissioning the device. • "Post-installation check" checklist see Section 3.3 • "Post-connection check" checklist see Section 4.4 Basic settings...
  • Page 16: Output Setting

    Commissioning Ceraphant T PTC31, PTP31, PTP35 Base Basic settings Note! The procentage specification applies to the upper range limit. Output setting • Hysteresis function: The hysteresis function enables two-point control via a hysteresis. Depending on the pressure p, the hysteresis can be set via the switch point SP and the switch-back point RSP.
  • Page 17 Ceraphant T PTC31, PTP31, PTP35 Commissioning OUT/OUT2 Output/output 2 Switching characteristic HYNO: Hysteresis/NO contact HYNC: Hysteresis/NC contact WINO: Window/NO contact WINC: Window/NC contact 4...20 mA: Analog output (only if available) Switch point value Switch point 0.5...100 % in increments of 0.1 % of selected unit (min.
  • Page 18: Service Function Setting

    Commissioning Ceraphant T PTC31, PTP31, PTP35 4 - 20 Analog output Value for 4 mA Enter lower range value in (LRV) increments of 0.1 % of selected unit Value for 20 mA Enter upper range value in (URV) increments of 0.1 % of...
  • Page 19: Maintenance

    Ceraphant T PTC31, PTP31, PTP35 Maintenance SERV Service functions Simulation OFF: No simulation output 1 or 2 OPEN: Switch output open CLOS: Switch output closed 3.5: Simulation values for analog output in mA (3.5/4.0/8.0/12.0/16.0/ 20.0/21.7) Max. indicator Display of max. measured process value Min.
  • Page 20: Accessories

    Accessories Ceraphant T PTC31, PTP31, PTP35 Accessories Process connection • Ceraphant T PTC31: The sensor module and the process connection are connected together and cannot be separated. • Ceraphant T PTP31, PTP35: The process connection is an adapter and the sensor module has an adapter thread. As a result, the process connection can easily be changed at a later stage.
  • Page 21 Ceraphant T PTC31, PTP31, PTP35 Accessories 8.1.1 Adapter change The adapter can be changed on PTP31 and PTP35. PTP31 PTP35 P01-PTx3xxxx-17-xx-xx-xx-001 Fig. 13: Changing the adapter ➀ Sensor module with adapter thread ➁ Standard O-ring ➂ Adapter Please note the following when changing the adapter: •...
  • Page 22 Accessories Ceraphant T PTC31, PTP31, PTP35 8.1.2 Adapter versions • PTP31: order numbers for thread adapter versions. Version AC: order no. 52023980 Version AD: order no. 52023981 Version AE: order no. 52023982 Version AF: order no. 52023983 Version BA: order no. 52023984 Version CA: order no.
  • Page 23 Ceraphant T PTC31, PTP31, PTP35 Accessories • PTP35: order numbers for Clamp adapter versions. SW28 AF28 Version DB: order no. 52023994 Version DL: order no. 52023995 Ø43.5 (1.7) Ø50.5 (2.0) Optional with inspection certificate 3.1: SW32 Version DB: order no. 52024001 AF32 Version DL: order no.
  • Page 24: Welding Bosses

    Accessories Ceraphant T PTC31, PTP31, PTP35 Welding bosses 8.2.1 Welding boss with sealing taper • Welding boss for flush mounting process connection G1A with metallic sealing taper (version BA for PTP35) Material: AISI 316L Order number: 52005087 ø50 (1,97) • Optional with inspection certificate 3.1...
  • Page 25: Electrical Connection

    Ceraphant T PTC31, PTP31, PTP35 Accessories Electrical connection 8.3.1 Plug-in jack, connecting cable • M12x1 plug-in jack. Connection to M12x1 housing connector Materials: Body PA, coupling nut Cu Zn, brass, nickeled. Protection: IP 67 (fully locked). P01-PMP13xxx-00-xx-00-xx-003 Order number: 52006263 •...
  • Page 26: Readwin

    Accessories Ceraphant T PTC31, PTP31, PTP35 ReadWin • Configuration kit. For PC–programmable transmitters. Setup program and interface cable for PCs with USB port. Adapter for transmitters with 4–pin post connector. Order code: TXU10-AA ® • ReadWin 2000: Supplied with the configuration kit or downloaded from internet: www.readwin2000.com...
  • Page 27: Trouble-Shooting

    Ceraphant T PTC31, PTP31, PTP35 Trouble-shooting Trouble-shooting Errors and warnings If an error in the device occurs, the colour of the status LED changes from green to red and the background illumination of the digital display changes from white to red. The display shows: •...
  • Page 28: Repair

    A repair is not planned. Return The following procedures must be carried out before a device is returned to Endress+Hauser: • Always enclose a fully completed “Declaration of Contamination” form with the device. Only then can Endress+Hauser transport and examine a returned device. A copy of the “Declaration of Contamination”...
  • Page 29: Change Status - History

    Ceraphant T PTC31, PTP31, PTP35 Technical data Change status - history Release no. device Changes 1.00 1.01 New analog electronics 1.02 Modification sensor units 1.03 Internal device modification 1.04 Internal device modification Technical data For technical data, please refer to the Technical Information Ceraphant T TI384P.
  • Page 30: Index

    Index Ceraphant T PTC31, PTP31, PTP35 Index Adapter change ......21 Adapter versions ......22 Additional operating options .
  • Page 31 Erklärung zur Kontamination und Reinigung Erklärung zur Kontamination und Reinigung Please reference the Return Authorization Number (RA#), obtained from Endress+Hauser, on all paperwork and mark the RA# Please reference the Return Authorization Number (RA#), obtained from Endress+Hauser, on all paperwork and mark the RA# clearly on the outside of the box.
  • Page 32 www.endress.com/worldwide KA225P/00/EN/10.09 71102965 FM+SGML 6.0 71102965...

This manual is also suitable for:

Ceraphant t ptp31Ceraphant t ptp35

Table of Contents