March 30, 2020
Addendum – Procedures for Converting Drive Gas from Oxygen to Air
Dear Valued Customer:
This communication is an addendum to the document noted above and provided on March 23, 2020 regarding
the use of Mindray A-Series Anesthesia Delivery Systems for continuous ventilation. Specifically detailed is the
process for converting the systems drive gas from oxygen to air.
Note: Only individuals qualified to perform maintenance/repair on anesthesia systems should carry
out this conversion.
Systems having the drive gas modified should be prominently labeled as such and may not be
used for anesthesia delivery, until the drive gas is reverted to oxygen.
Guidance for converting the A7 system is provided under Attachment 1.
Guidance for converting the A5 system is provided under Attachment 2.
Guidance for converting the A3/A4 system is provided under Attachment 3.
The following are recommendations/considerations during use after a system's drive gas has been converted
from oxygen to air:
1. It is strongly recommended to connect both the air and oxygen source before using the system to achieve
a gas source pressure between 280-600 kPa (41-80 psi).
2. If the pressure of air is lower than 140 kPa ±10% (20 psi ±10%), the system will display the alarm "Drive
Gas Pressure Low" and will not ventilate.
3. If the pressure of the oxygen gas source is lower than 220 kPa ±10% (32 psi ±10%) the system will display
the alarm "O2 Supply Failure". Under this condition, the system will continue to ventilate but the oxygen
fresh gas delivered may be lower than expected.
4. If the flow sensor requires calibration after the drive gas has been converted, the tidal volume accuracy may
be up to 5% over specification.
Mindray's in-house supports team remain available 24/7. If assistance is needed, please contact Technical
Support at (800) 288-2121, option 2.
Sincerely,
Diane Arpino
Director, Quality Operations and Regulatory Affairs
Mindray DS USA, Inc.
US Agent for Shenzhen Mindray Biomedical Corp., Ltd.
d.arpino@mindray.com
Mindray North America
Mindray A-Series Anesthesia Delivery System
Consideration for use as a Ventilator
800 MacArthur Blvd.
Mahwah, NJ 07430 Tel: 201.995.8000
Page 1 of 19
www.mindray.com
0002-08-1659 Rev. A
Need help?
Do you have a question about the A Series and is the answer not in the manual?
Questions and answers