If a sensor option is enabled on the LCD but the new sensor
is not installed on the detector, the following message
displays: H2S SEnSor not inStALLEd - SEnSor CAnnot bE
EnAbLEd.
3.
Use G or H to scroll to a new user option, or press
C to exit user options and return to normal
operating mode.

Pass Code Protection

The PASS Lock option is used to prevent access to the user
options and calibration/alarm setpoint procedure without a
pass code.
For calibration, the pass code is requested after
auto zero is performed (if enabled).
If the detector is pass code protected, the key icon (
displays at the bottom of the LCD during start-up.
The detector is shipped with the pass code
protection option disabled.
Enable/Disable Pass Code Protection
To enable/disable the pass code protect option, complete the
following:
Note
)
Note
1.
From the user options menu, use G or H to scroll
to the PASS Lock option.
2.
Press C to select this option.
The detector emits two beeps and the LCD displays
if pass code protection is enabled or disabled.
or
The LCD displays the current mode (On or OFF)
and then returns to the FiniSh OPtionS screen.
Note
The pass code is provided on a separate card
inside the shipping container.
GasAlertMicro
User Options Menu
25