When each of the span concentration values have
been accepted, the LCD returns to the FiniSh
OPtionS screen.
6.
Use G or H to scroll to a new user option, or press
C to exit user options and return to normal
operating mode.

STELCalculation Period

The StEL PEriod option is used to adjust the short-term
exposure limit (STEL) calculation period (5-15 minutes).
The detector is shipped with the STEL calculation
period set to 15 minutes.
To change the STEL calculation period, complete the
following:
1.
From the user options menu, use G or H to scroll
to the StEL PEriod screen.
Note
Press C to select StEL PEriod and to access the STEL time
period screen.
2.
Use G or H to scroll to the required time period
value (5-15). Press C within 10 seconds to save
the new period change.
The LCD returns to the FiniSh OPtionS screen.
If the new value is not entered and confirmed within
10 seconds, the LCD displays an error message.
3.
Use G or H to scroll to a new user option, or press
C to exit user options and return to normal
operating mode.
GasAlertMicro
User Options Menu
Note
29