Setting the Remaining Alarm Setpoints
12. Repeat steps 8-10 to set the alarm setpoints for
each remaining sensor.
The audible alarm beeps four times when the alarm
setpoint function is complete. The detector then
proceeds to the CAL duE screen.
Setting the Calibration Due Date
The calibration due date is designed to automatically display
the number of days remaining before calibration is due to be
performed.
The CAL duE screen automatically displays how many days
are remaining during the self-test process. The calibration
due date option allows from 1-365 days to be set as the
default value. BW recommends that the detector be
calibrated at least once every 180 days (6 months).
Calibration and Setting Alarm Setpoints
The detector is shipped with the factory default
setting of 180 days.
Successful Calibration
When the sensors are successfully calibrated, the calibration
due date automatically resets to the previous entered setting.
To change the default number of days remaining before
calibration is due, complete the following:
13. From the CAL due screen, use G or H to scroll to
a new value (1-365).
14. Press C to confirm the new value. The new value
becomes the default value. The detector then
beeps five times and returns to normal operation.
If a new value is not confirmed within 10 seconds,
the detector automatically retains the previous
calibration due date.
GasAlertMicro
Note
Note