A&D Life Source UA-767FAM Instruction Manual page 7

Table of Contents

Advertisement

Available languages

Available languages

Symbols
Appears during measurement when
the cuff is attached loosely
Detected rate of IHB in memory
User
Previous measurements stored in memory.
Average data
The battery is full, and is shown
during measurement.
The battery is low when it blinks.
Unstable blood pressure due to
movement during measurement.
The systolic and diastolic values are
within 10 mmHg of each other.
The pressure value did not increase
during the inflation.
The cuff is not applied correctly.
PUL DISPLAY ERROR
The pulse was not detected correctly.
Blood pressure monitor internal error
Time in the morning
AM
PM
Time in the afternoon.
Symbols
Function / Meaning
English 6
Recommended Action
The reading may yield an
incorrect value. Apply the
cuff correctly and take
another measurement.
Replace all batteries with new
ones when the mark blinks.
Take another measurement.
Remain very still during
measurement.
Apply the cuff correctly and
take another measurement.
Remove the batteries and
press the
button,
START
and then install the batteries
again. If the error still appears
contact the manufacturer.