This addendum is to be used in conjunction with the Owners Manual for the F-2000 Hay Moisture Meter
(attached). This addendum contains information specific to the tobacco applications. We recommend the user
read the Owner's Manual and addendum completely, before using the F-2000T.
The meter reads moisture in TOBACCO between 8.0 and 37.0%. It also stores the reading and
displays the average and maximum of up to 100 accumulated readings. The meter also has a set
point and a buzzer. Whenever the displayed reading is above the set point value, the buzzer will
sound. The meter will automatically correct the input reading for sample temperature, provided that
the user has entered the correct temperature in the meter.
Whenever the battery is removed and then reconnected (i.e when power is applied), the meter will
display the software version number (4.2) for one second and then shut off if no key is pressed.
The meter has FOUR keys - the "READ" key, the "CHECK" key, the "SET" key and the "TEMP" key
(see the attached diagram).
1. TO RESET THE METER
Resetting the meter will restore the default settings and erase all the stored readings.
Press the "CHECK" key immediately, followed by the "SET" key. The meter will display "200" for two
seconds and then shut off.
2. TO CHANGE THE SET POINT VALUE
The default set point value for the meter is 20.0%. The user can change the value between 8.0 and
37.0.
Press the "SET" key. The meter will display the current set point value. To increase the set point,
keep the "SET" key pressed. The value will increase in steps of 1.0. To decrease the value, press the
"TEMP" key within two seconds after pressing the "SET" key. The value will decrease in steps of 1.0.
3. TO CHANGE THE TEMPERATURE SETTING
The meter will automatically correct the moisture reading for sample temperature, provided the user
has entered the correct SAMPLE TEMPERATURE in the meter. The user can set the temperature
between 0º and 255º F.
The default temperature setting is 80º F. The meter will correct the moisture value by 0.05% for every
degree change in the sample temperature.
Press the "TEMP" key. The meter will display the current temperature setting. To increase the setting,
keep the "TEMP" key pressed. The temperature value will increase in steps of 5º. To decrease the
value, press the "SET" key within two seconds after pressing the "TEMP" key. The value will decrease
in steps of 5.0.
OWNERS MANUAL FOR F-2000T
(VERSION 4.2 FOR BURLEY TOBACCO)
Tel: +1 (281) 516-9292 / (888) 275-5772 eFax: +1 (866) 234-0451
Web: https://www.abqindustrial.net E-mail: info@abqindustrial.net
Distributed by: ABQ Industrial LP USA

Advertisement

loading
Need help?

Need help?

Do you have a question about the F-2000T and is the answer not in the manual?

Questions and answers

Summary of Contents for Delmhorst F-2000T

  • Page 1 (attached). This addendum contains information specific to the tobacco applications. We recommend the user read the Owner’s Manual and addendum completely, before using the F-2000T. The meter reads moisture in TOBACCO between 8.0 and 37.0%. It also stores the reading and displays the average and maximum of up to 100 accumulated readings.
  • Page 2 4. READING THE %MC VALUE Press the “READ” key. Within 1 or 2 seconds, the meter will display the %MC between 8.0 and 37.0. The meter will automatically correct the reading for the sample temperature. The reading will be displayed for 2 seconds and if the “READ” key is still pressed, then the cycle will be repeated. The buzzer will sound if the displayed reading is above the set point value entered in the meter (default is 20.0).
  • Page 3 Connector For external electrode Digital Readout Temperature Button (TEMP) %MC Read Button (READ) Set Point Button (SET) Calibration Check Button (CHECK)
  • Page 4 TABLE OF CONTENTS F-2000 Features Belo re Vou Begi n Check Calibration Set the Sca l e Change the Set-Point To Check Accumulated Readings To Reset Meter Taking a Reading Testing Baled Hay Testing in the Windrow F actors Affecti ng Vou r Readi ng Range of Moisture Content Hay Temperature Curing...
  • Page 5: Change The Set-Point

    CHANGE THE SET-POINT@ ► To change lhe set-polnt value press lhe set-polnl button The meter wlll dlsplay the current set-polnt value. ► To scroll foiward to a hlgher value, hold the set-polnt button8 v. f llle the current value Is dlsplayed and scroll to the set-polnt value deslred.
  • Page 6 Distributed by: ABQ Industrial LP USA Tel: +1 (281) 516-9292 / (888) 275-5772 eFax: +1 (866) 234-0451 Web: https://www.abqindustrial.net E-mail: info@abqindustrial.net...
  • Page 8 Density The callbratlon of the molsture testers applles to bales of normal ''average' denslty. Generally: ► Oenser bales may yleld readings 1-2% polnts hlgher. ► Looser bales tend to yleld 1-2% polnt lower. ► Tests ln stacks usually yleld readlngs 2%-3% lower. ►...
  • Page 9 Distributed by: ABQ Industrial LP USA Tel: +1 (281) 516-9292 / (888) 275-5772 eFax: +1 (866) 234-0451 Web: https://www.abqindustrial.net E-mail: info@abqindustrial.net...
  • Page 10 Distributed by: ABQ Industrial LP USA Tel: +1 (281) 516-9292 / (888) 275-5772 eFax: +1 (866) 234-0451 Web: https://www.abqindustrial.net E-mail: info@abqindustrial.net...