The front number represents the number of verification record
document, and the rear number represents the pressure point number to
be recorded. When the measurement of a certain pressure point is
completed, just long press the "
automatically save the current measured data, and the pressure point
number on the "verified data record identification" will increase by one
digit, waiting for the measurement of the next pressure point. When all
the pressure points to be measured are measured and recorded, the
verified data recording will be completed, and the "verified data
recording identification" on the upper part of the interface will
automatically disappear.
8.3

Verification completed

As mentioned above, when all the pressure points to be measured
are measured and recorded, the verified data recording will be
completed, and the "verified data recording identification" at the upper
part of the interface will automatically disappear. However, if the
pressure point data is not fully recorded, then the verification recording
process can be ended as follows:
↓ "
key and the calibrator will
16