Stellar basics

About the control panel

Key
Start/Stop
Alarm mute
Menu keys
Push dial
Function
Starts or stops treatment.
Extended hold for at least three seconds starts the mask-fitting
feature.
During therapy: Press once to mute an alarm. Press a second time to
un-mute an alarm. If the problem is still present, the alarm will sound
again after two minutes. Refer to Working with alarms (see page 16)
In stand-by mode: Extended hold for at least three seconds starts the
alarm LED and alarm buzzer test.
Press the appropriate menu button (Monitoring, Setup, Info) to enter and
scroll through the relevant menu.
Turning the dial allows you to scroll through the menu and change
settings. Pushing the dial allows you to enter into a menu or confirm
your choice.