General dangers of failing to follow the safety instructions The Z30, Z30–G force transducer corresponds to the state of the art and is fail-safe. The transducers can give rise to remaining dangers if they are inappropriately installed and operated by untrained personnel.
Page 5
Meaning: CE mark The CE mark signals a guarantee by the manufacturer that his product meets the requirements of the relevant EC directives (see Declaration of conformity at the end of this document). HBM 23.8.00 SUNSTAR自动化 http://www.sensor-ic.com/ TEL: 0755-83376489 FAX:0755-83376182 E-MAIL:szss20@163.com...
Page 6
Protect the transducer from damp and weather influences such as rain, snow, etc. Maintenance The Z30, Z30–G force transducer is maintenance free. Accident prevention Although the specified nominal force in the destructive range is several times the full scale value, the relevant accident prevention regulations from the trade associations must be taken into consideration.
Order number: D–MS/MONT 15–pin D–Sub connector (male) fitted to Kab139A; Order no.: D–15D/MONT Z30–G is only available in conjunction with DKD calibration carried out by HBM according to EN10002–3. The calibration certificate must be ordered separately (see accessories). HBM 23.8.00...
SUNSTAR传感与控制 http://www.sensor-ic.com/ TEL:0755-83376549 FAX:0755-83376182 E-MAIL:szss20@163.com Application notes Force transducer Z30: Series Z30 precision force transducers measure tensile and compressive forces. They measure static and quasi-static forces with great accuracy and reproducibility and therefore require judicious handling. You must be particularly vigilant when transporting and installing the devices. If the transducers are dropped or jolted, permanent damage could be caused.
3.3 Disturbance variables Torsion, bending and transverse load are disturbance variables and are therefore to be avoided. If necessary they can be remedied with HBM mounting accessories (chapter 5.3 ). The effects of temperature on the zero signal (S.G. bridge and housing) and on the sensitivity are compensated.
The aluminium enclosures of the transducers is protected by a powder coating. If used in difficult environmental conditions (direct weather effects, contact with media which encourage corrosion) additional protective measures should be employed by the user. HBM 23.8.00 SUNSTAR自动化 http://www.sensor-ic.com/ TEL: 0755-83376489 FAX:0755-83376182 E-MAIL:szss20@163.com...
The measurement direction in which forces act on the transducer must be as precise as possible. WARNING Torsion and bending moments, eccentric loading and transverse forces result in measurement errors and if limit values are exceeded, could destroy the transducer. HBM 23.8.00 SUNSTAR自动化 http://www.sensor-ic.com/ TEL: 0755-83376489 FAX:0755-83376182 E-MAIL:szss20@163.com...
SUNSTAR传感与控制 http://www.sensor-ic.com/ TEL:0755-83376549 FAX:0755-83376182 E-MAIL:szss20@163.com 5.3 Mounting accessories for tensile loading/compressive loading Knuckle eyes are available for installing with the Z30. These mounting accessories prevent the introduction of torsional moments into the transducer. Using two knuckle eyes will also exclude bending moments, shear loading and eccentric loading.
SUNSTAR传感与控制 http://www.sensor-ic.com/ TEL:0755-83376549 FAX:0755-83376182 E-MAIL:szss20@163.com Electrical connection 6.1 Instructions for cabling Always use shielded, low-capacity measurement cable (HBM cables meet these requirements). Do not lay measurement cable parallel to high-voltage power lines or control circuits. If this is not possible (e.g. in cable ducts) protect the measurement cable, e.g.
Page 14
Excitation voltage (+) U GN (green) Sensor circuit (+) GY (grey) Sensor circuit (–) shielding Cable shielding, connected to housing Fig. 6.2: Pin assignment on the unterminated Z30/Z30-G (for KAB139A–6 connection cable) HBM 23.8.00 SUNSTAR自动化 http://www.sensor-ic.com/ TEL: 0755-83376489 FAX:0755-83376182 E-MAIL:szss20@163.com...
–10...+70 Storage temperature range –25...+85 Reference temperature Max. operational force Limit force Breaking force Static lateral limit force See supplementary information according to DIN EN10002–3 resp. ISO378 for Z30–G on following page HBM 23.8.00 SUNSTAR自动化 http://www.sensor-ic.com/ TEL: 0755-83376489 FAX:0755-83376182 E-MAIL:szss20@163.com...
Page 16
Effect of temperature on 0.02 0.02 (typ.0.01) zero signal/10K by ref. to sensitivity Z30–G is only available in conjunction with DKD calibration carried out by HBM according to EN10002–3. HBM 23.8.00 SUNSTAR自动化 http://www.sensor-ic.com/ TEL: 0755-83376489 FAX:0755-83376182 E-MAIL:szss20@163.com...
SUNSTAR传感与控制 http://www.sensor-ic.com/ TEL:0755-83376549 FAX:0755-83376182 E-MAIL:szss20@163.com Dimensions Z30, Z30–G (50 N – 1000 N) Male connector with protecting cap Male connector a.f. series 723, 7–pin HBM 23.8.00 SUNSTAR自动化 http://www.sensor-ic.com/ TEL: 0755-83376489 FAX:0755-83376182 E-MAIL:szss20@163.com...
Page 18
SUNSTAR传感与控制 http://www.sensor-ic.com/ TEL:0755-83376549 FAX:0755-83376182 E-MAIL:szss20@163.com Dimensions Z30, Z30–G (2 kN – 10 kN) (continued) Male connector series 723, 7-pin Angle variable according to mounting Male connector, series 723, 7-pin, with protecting a.f. HBM 23.8.00 SUNSTAR自动化 http://www.sensor-ic.com/ TEL: 0755-83376489 FAX:0755-83376182 E-MAIL:szss20@163.com...
Page 19
Knuckle eye ZGOW for nominal load 50 N to 1000 N Lock nuts included in the scope of supply Type U1R/200kg/ZGW * Centre-to-centre distance in assembled state (transducer with knuckle eyes above and below) HBM 23.8.00 SUNSTAR自动化 http://www.sensor-ic.com/ TEL: 0755-83376489 FAX:0755-83376182 E-MAIL:szss20@163.com...
Page 20
Nominal load 50 N – 1 kN Nominal load 2 kN – 10 kN Ball R Ball R Type U1R/200kg/ZL Nominal force Weight (kg) 50 N ... 1 kN 2 kN ... 10 kN 0.026 HBM 23.8.00 SUNSTAR自动化 http://www.sensor-ic.com/ TEL: 0755-83376489 FAX:0755-83376182 E-MAIL:szss20@163.com...
Page 21
EN10002–3 or ISO376 calibrations –0.1 groove 5 x 0.5 deep α Weight Type Thrust piece (kg) approx. Z30/ 50 N...1000 N EDO3/1 kN 13.2 Z30/ 2 kN...10 kN EDO4/50 kN 0.34 20.2 HBM 23.8.00 SUNSTAR自动化 http://www.sensor-ic.com/ TEL: 0755-83376489 FAX:0755-83376182 E-MAIL:szss20@163.com...
Page 24
Modifications reserved. OTTINGER ALDWIN ESSTECHNIK All details describe our products in general HBM Mess- und Systemtechnik GmbH form only. Postfach 10 01 51, D-64201 Darmstadt They are not to be understood as express Im Tiefen See 45, D-64293 Darmstadt warranty and do not constitute any liability Tel.: +49/ 61 51/ 8 03-0;...