5.3 Connection of the power adapter

1)
Connect the power adapter to the device's
power cord.
2)
Connect the power adapter to the wall outlet.

5.4 Therapy Mode

Sonic Stim + offers 2 treatment modes:
1)
Combination: Ultrasound + Electrical Stimulation therapy
2)
Ultrasound: Ultrasound therapy

5.5 Disconnecting

1)
Turn off the Sonic Stim + by switching power ON/OFF switch from ON to
OFF position.
2)
Remove the power adapter from the wall outlet.

6.0 Operation

6.1. Measures with regards to treatments

Before treatment
Ensure there are no contraindications to treatment.
Clean the skin of the treatment area with soap or alcohol (70%).
If the skin has excess hair, trim or shave hair for optimal treatment.
Apply a reasonable amount of ultrasound transmission/conductive gel to
the treatment area. Use only the ultrasound gel with CE mark.
Ultrasonic Action Function Test: Place the probe horizontally, then apply
several water drops on the surface of the probe, turn the device on and
press the time button to activate the ultrasound device. You will be able
to observe the ultrasonic action as the water droplets will appear to be
dancing on the sound head and you may notice a small amount of
steam being released. The water droplets on the probe start to perform
one million vibrations per second showing the atomization phenomenon.
During treatment
Move the ultrasound-head in a circular motion. The area treated should
be twice the size of the diameter of the applicator.