Summary of Contents for RKI Instruments 35-3001A-05 Series
Page 1
35-3001A-05-XX Carbon Dioxide Sample-Draw Detector Operator’s Manual Part Number: 71-0340 Revision: B Released: 6/10/19 www.rkiinstruments.com...
Page 2
WARNING Read and understand this instruction manual before operating detector. Improper use of the detector could result in bodily harm or death. Periodic calibration and maintenance of the detector is essential for proper operation and correct readings. Please calibrate and maintain this detector regularly! Frequency of calibration depends upon the type of use you have and the sensor types.
Page 3
Product Warranty RKI Instruments, Inc. warrants gas alarm equipment sold by us to be free from defects in materials, workmanship, and performance for a period of one year from date of shipment from RKI Instruments, Inc. Any parts found defective within that period will be repaired or replaced, at our option, free of charge. This warranty does not apply to those items which by their nature are subject to deterioration or consumption in normal service, and which must be cleaned, repaired, or replaced on a routine basis.
Overview This operator’s manual describes the 35-3001A-05-XX carbon dioxide sample-draw detector. This manual also describes how to install, start up, maintain, and calibrate the sample-draw detector when using it with a gas monitoring controller. A parts list at the end of this manual lists replacement parts and accessories for the sample-draw detector.
Description This section describes the components of the carbon dioxide sample-draw detector. Pump Connec tor Mounting Foot, 4X Flowmeter Circuit IR Amplifier Board Connector Ox ygen Sensor Terminal Sensor Flow Senso r Strip Cur ren t Control Valv e 14 8mA Pump Fail LED Pilot LED...
External Components This section describes the sample-draw detector’s external components. Housing The sample-draw detector’s fiberglass housing is weather- and corrosion-resistant. It is suitable for installation where general purpose equipment is in use. The housing door is hinged on the left side and is secured by two latches on the right side. The flowmeter and status LEDs are visible through a window in the housing door.
Page 8
Main Circuit Board The main circuit board includes the detector/amp terminal strip, interconnect terminal strip, oxygen sensor terminal strip, pump connector, and flowmeter circuit board connector (see Figure 1). Detector/Amp Terminal Strip The detector/amp terminal strip is the upper twelve-point terminal strip in the bottom right corner of the main circuit board.
Page 9
Status LEDs Two status LEDs are above the flowmeter. They are also visible through the window in the housing door. The green Pilot LED is on when the sample-draw detector is receiving power from the controller. The red Fail LED is on when the sample flow rate is below the low flow level. Pressure Switch The pressure switch is mounted to the back of the flowmeter circuit board.
Page 10
Amplifier The amplifier converts the electrical output from the sensor to a 4 to 20 mA signal that corresponds to the detection range and transmits the signal to a gas monitoring controller. A label on the amplifier indicates the detector drive current. This drive current is factory set. The amplifier is mounted on the top middle edge of the main circuit board.
Installation This section describes procedures to mount the sample-draw detector in the monitoring environment and wire the sample-draw detector to a controller. Mounting the Carbon Dioxide Sample-Draw Detector Select the mounting site. Consider the following when you select the mounting site: •...
Connect a length of sample tubing to the other side of the particle filter and route it to the sampling area. 1/4” O.D. rigid polypropylene, Teflon, or flexible polyurethane tubing may be used. RKI Instruments, Inc. recommends using either 1/4” O.D. x 1/8” I.D. or 1/4” O.D. x 0.170” I.D. tubing based on your length requirements.
Push 1/4” O.D. rigid polypropylene or rigid Teflon sample tubing into the fitting until it stops. Flexible polyurethane tubing may be used with an appropriate insert. RKI Instruments, Inc. recommends using either 1/4” O.D. x 1/8” I.D. or 1/4” O.D. x 0.170” I.D. tubing based on your length requirements. See “Specifications”...
Page 14
Connect the wires to the applicable detector/transmitter terminal strip at the controller as shown in Figure 5. Refer to the controller operator’s manual and the controller detector head specification sheet for the 35-3001A-05-XX for detector/terminal strip connections specific to the controller. Not Used on T his V ersion Z E RO...
Start Up This section describes procedures to start up the sample-draw detector and place the sample-draw detector into normal operation. Introducing Incoming Power Complete the installation procedures described earlier in this manual. Verify that the wiring is correct and secure. Refer to the controller operator’s manual for connections at the controller.
Setting the Zero Reading Since there is a background of CO in air of typically 300 - 600 ppm (0.03 - 0.06% volume), it is necessary to use a calibration kit with a 100% nitrogen cylinder to set the zero signal for the 0 - 5,000 ppm and 0 - 5% volume versions.
If the fail condition continues, replace the sensor as described in “Replacing the Infrared Carbon Dioxide Sensor” on page 18. If the fail condition continues, contact RKI Instruments, Inc. for further instruction. Slow or No Response/Difficult or Unable to Calibrate Symptoms •...
If the calibration/response difficulties continue, contact RKI Instruments, Inc. for further instruction. Replacing Components of the Carbon Dioxide Sample-Draw Detector This section includes procedures to replace the sensor, hydrophobic filter, and particle filter. Replacing the Infrared Carbon Dioxide Sensor Turn off the controller.
Page 19
Install the detector and controller plug-in terminals strips into their sockets on the new amplifier. If controller leads or detector cable leads were removed from the plug-in terminal strips during this procedure, refer to Table 2 and Table 3 below. Table 2:Reconnecting the Amplifier to the Detector/Amp Terminal Strip Amplifier Controller...
Replacing the Particle Filter Turn off the controller. Turn off power to the controller. If the particle filter is installed at the inlet fitting: Disconnect the tubing routed to the sampling area, if installed, from the particle filter. Disconnect the particle filter from the tubing stub. Install the new particle filter onto the tubing stub.
Removing the Particle Filter’s Tubing Stub, if Necessary A short tubing stub comes factory installed in the particle filter. It is used for connecting the particle filter to the inlet fitting. If you have installed the particle filter and no longer want it installed, you will need to remove the particle filter’s tubing stub from the inlet fitting and replace it with tubing.
Although there is no particular calibration frequency that is correct for all applications, a calibration frequency of every 6 months is adequate for most infrared carbon dioxide sample draw applications. Unless experience in a particular application dictates otherwise, RKI Instruments, Inc. recommends a calibration frequency of every 6 months.
Unscrew the regulator from the 100% nitrogen calibration cylinder. Setting the Response Reading Screw the regulator into the carbon dioxide calibration cylinder. Connect the sample tubing from the demand flow regulator to the sample-draw detector’s inlet line. Allow the sample-draw detector to draw sample for one minute. When the reading on the voltmeter stabilizes, verify that the reading matches the response reading (±2 mV) you determined earlier.
Parts List Table 4 lists replacement parts and accessories for the carbon dioxide sample-draw detector. Table 4: Parts List Part Number Description 06-1248RK Sample tubing, 3/16 in. ID x 5/16 in. OD, specify length 06-1248RK-03 Sample tubing, 3/16 in. ID x 5/16 in. OD, 3 feet (for calibration kit) 07-0110RK Gasket for IR flow block 17-2670...
Need help?
Do you have a question about the 35-3001A-05 Series and is the answer not in the manual?
Questions and answers