Troubleshooting Guide

° Check that all the pump components have been assembled correctly and that the
white valve is positioned correctly.
° Ensure that the silicone diaphragm and Let-down Massage Cushion are fitted
Lack of suction
securely to the pump body and a perfect seal is created, making sure to push the
Let-down Massage Cushion far enough down the center of the funnel.
° Make certain that the tubing connections and control handle are firmly in place.
° Ensure the pump is correctly assembled and that suction is being created.
No milk being
° Relax and try again, as expression may improve after practice.
expressed
° Refer to the 'Preparing to Pump' section on page 7.
° You may be pumping too hard.You do not need to use all the suction the pump
Pain in the
can generate.Try compressing the handle halfway.
breast area
° Try holding down the pump handle for 2-3 seconds per compression.
when expressing
° Consult your health professional or breastfeeding advisor.
° Try leaning slightly forward.
° Remove and re-fit the Let-down Massage Cushion to ensure it is firmly
Milk is drawn up
attached and pushed far enough down the center of the funnel.
the Let-down
Some women may prefer to use the pump without the Let-down Massage Cushion. The Let-down
Massage Cushion
Massage Cushion has petals that flex in and out to gently massage the area around the breast.
This is intended to help stimulate a fast, natural milk flow. When using the breast pump without
the Let-down Massage Cushion, the maximum vacuum level achievable will be higher.
° Avoid contact with antibacterial or abrasive detergents or cleaners as they
can damage the plastic. Combinations of detergents, cleaning products, sanitizing
Cracking or
solution, softened water and temperature fluctuations may, under certain
discoloration
circumstances, cause the plastic to crack. If this occurs, do not use. The pump
bodies and breast milk containers are dishwasher safe but food colorings may
of pump parts
discolor them. Follow guidelines in 'Cleaning and sanitizing your
Your Pump' on page 6.
° Contact Philips AVENT for replacements.
Lost or broken parts
Call 1.800.54.AVENT or visit www.philips.com/AVENT
° Ensure the pump is correctly assembled and switched on with the power indicator
Pump not responding
illuminated GREEN.
to pressing of the
° Make sure the cables between the electric wall socket and the pump are correctly
control handle
connected.
° If the pump stops working and the power light is flashing
detected.
° 3 flashes, delay, 3 flashes, etc – the breast pump has detected a vacuum fault. Switch
off the motor and check tubing and connections for obstructions or damage.
Pump light
° 2 flashes, delay, 2 flashes, etc – the breast pump has detected that it is over the
flashing
RED
acceptable temperature. Switch off the motor. Ensure unit is not exposed to heat
sources or sitting in direct sunlight. Allow unit to stabilize before trying again.
° Continuous flashing – the pump has detected that the control cable has been
incorrectly fitted/not fitted or the cable is damaged.
21

Technical Specifications

° Mains Input Voltage 100 – 240 VAC
° Mains Input Current 1.5A
° Mains Input Frequency 50 – 60 Hz
° Supply voltage fluctuations not exceeding
FULL ONE-YEAR WARRANTY
Philips Electronics North America Corporation (USA) and Philips Electronics Ltd (CANADA)
warrant each new Philips AVENT product, model SCF304, against defects in materials or
workmanship for a period of one year from the date of purchase, and agrees to repair or
replace any defective product without charge. IMPORTANT:This warranty does not cover
damage resulting from accident, misuse or abuse, lack of reasonable care, or the affixing of any
attachment not provided with the product. NO RESPONSIBILITY IS ASSUMED FOR ANY
SPECIAL, INCIDENTAL OR CONSEQUENTIAL DAMAGES.
In order to obtain warranty service, simply call toll-free 1.800.54.AVENT.
RED
then a fault has been
Need Help? Call us at 1.800.54.AVENT or visit www.philips.com/AVENT
Key to product base label
TYPE B APPLIED PART
ETL LISTED MARK –
CANADA/UNITED STATES
LISTED
3045297
° Ambient temperature range: +10° C to
+40° C (+50° F to +104° F)
° Relative Humidity range: 30% to 75%
° Atmospheric Pressure: 700 to 1060hPa
+/- 10% of the nominal voltage
° Equipment rated for continuous operation