Additional Functions - Level Pro ULTRAPRO 1000 Instruction Manual

Ultrasonic level meters, reliable non-contact level
Table of Contents

Advertisement

11 .3 .  a
dditional
functions
Additional functions include modes to display temperature in the tank or to find out the actual flowing
current in the loop. Besides, to lock modifications using a password and information about the level
meter version. All of the functions are accessible from the main menu.
DIAGNOSTICS
It contains information about the actual temperature inside the tank
(or about the compensated temperature) "TEMPERATURE" and
current flow in through the loop "CURRENT". If the temperature
compensation ("MEDIUM TEMPERATURE") is activated, the cor-
rected temperature is displayed.
D I S T A N C E T O L E V E L
T E M P E R A T U R E
C U R R E N T
D I S T A N C E T O L E V E L
T E M P E R A T U R E
C U R R E N T
D I S T A N C E T O L E V E L
T E M P E R A T U R E
C U R R E N T
CLONE SETTINGS
This mode is intended for copying of the level meter (UltraPro
1000 body) configuration  into  the  display  module (DM–
100) and back. The display module can then be removed
from the level sensor body and put into another level
sensor transferring the settings (cloning).
The "CLONE SETTINGS" mode transfers all data, excluding setting of the "Teaching" and HART
C L O N E S E T T I N G S
S E N S O R
D I S P L A Y
M O D U L E
D I S P L A Y
S E N S O R
M O D U L E
1.
Press
to enter the menu and select the item "CLONE SETTINGS". Copying of the settings
from the body of the level meter to display module is done by selecting "SENSOR → DISPLAY
MODULE". To transfer the settings from the display module to another level meter select the item
DISPLAY MODULE → SENSOR.
2.
The selected mode starts by pressing button
CLONING" .
3.
After completing the process in the middle of the screen displays "DONE". It is then possible to
leave the menu and the mode by pressing the button
© Levelpro, s.r.o.
Ultrapro-1000
D I S T A N C E T O L E V E L
0 1 2 2 0 m m
The temperature is measured inside the tank
T E M P E R A T U R E
where the level meter is installed.
If the temperature of the measured medium is
+ 2 3 ° C
different, we recommend you to carry out the
temperature compensation "MEDIUM TEM-
PERATURE" because of accuracy (see p. 15).
Then the displayed temperature is an average
C U R R E N T
value from the temperature set in the "MEDIUM
TEMPERATURE" and the actual temperature
1 7 . 3 0 m A
measured by the sensor.
C L O N E S E T T I N G S
C L O N E S E T T I N G S
S E N S O R
D I S P L A Y
M O D U L E
D I S P L A Y
S E N S O R
M O D U L E
A R E Y O U S U R E ?
During transmission the display shows "NOW
.
B A S I C S E T T I N G S
S E R V I C E
D I A G N O S T I C
C L O N E S E T T I N G S
P A S S W O R D
L A N G U A G E
I N F O
B A S I C S E T T I N G S
S E R V I C E
D I A G N O S T I C
C L O N E S E T T I N G S
P A S S W O R D
I N F O
®
N O W C L O N I N G
.
20