Agilent 10400B Series Miniature Passive Oscilloscope Probes
Bandwidth considerations
Bandwidth considerations
The dominant probe limitation to system bandwidth is its input capacitance,
assuming that the high-frequency compensation adjustments have been made.
The displayed bandwidth of any measurement system using an oscilloscope and
probe is determined by four factors: probe input capacitance, source impedance,
source bandwidth, and oscilloscope bandwidth.
Agilent 10400B Series Probes Selection Guide
Model
Type of Probe Bandwidth
Number
10433B
High-
impedance,
passive
10436B
High-
impedance,
passive
10437B
50Ω Resistive
Divider
10439B
High-
impedance,
passive
10440B
High
impedance
passive
10441B
High-
impedance,
passive
10442B
500-Ω
resistive
divider
The Agilent 10437B and Agilent 10442B probes are compatible with
oscilloscopes that have a 50 Ω input.
Probe compatibility
Infiniium
Oscilloscope
8000A Series
10
Division
Scope/
Ratio
Probe
System
300 MHz
10:1
300 MHz
10:1
1 GHz
1:1
n/a
1:1
300 MHz
100:1
500 MHz
10:1
1 GHz
10:1
Compatible 10400B Series Probe
10433B, 10437B, 10439B, 10440B, 10442B
Input R Input C
10 MΩ 10 pF
10 MΩ 11 pF
50Ω
n/a
1 MΩ
n/a
10 MΩ 2.5 pF
1 MΩ
9 pF
500Ω
n/a
Scope
Comp
Length
Input R
Range
(in
meters)
1 MΩ
10-16 pF 2
1 MΩ
18-22 pF 2
50Ω
n/a
2
1 MΩ
n/a
1.5
1 MΩ
6-14 pF
2
1 MΩ
6-9 pF
1.8
50Ω
n/a
1.5