External Calibration/Adjustment with a Factory-Defined Weight
Externally calibrate weighing system and adjust using factory-defined (standard) weight
Step
isoTEST
1.
Select
function
2.
Select "Weighing system settings"
3.
Select external calibration/adjustment
with factory-defined weight
(tare the weighing system, if necessary)
4.
Start external calibration
5.
Place standard weight (e.g., 50 g)
on the weighing system
Minus sign –: weight too low
Plus sign +: weight too high
No plus/minus sign: weight o.k.
This is displayed after calibration:
(On verified analyzers, the deviation
between the readout and the conventional
measurement is displayed)
6.
If you do not need to adjust
the weighing system
52
Key (or instruction)
T key
soft key
O
soft key
O
Start
soft key
Load standard weight
End
soft key
Display/Printout
0000g
+
WGH.SYS: Ext. cal. factory-def. wt
oo
o
F
-
50000g
WGH.SYS: Ext. cal. factory-def. wt
Start
o
F
50000g
WGH.SYS: Ext. cal. factory-def. wt
Start
o
F
0001g
+
WGH.SYS: Ext. adj. factory-def. wt
End
o
--------------------
07.09.1999
13:03
Mod.
MA100C
Ser. no.
90805355
Ver. no.
01-38-07
ID
--------------------
External calibration
Nom.
+
50.000 g
Diff. +
0.001 g
--------------------
07.09.1999
13:03
Name:
--------------------
Start
Tare
Tara
q
Tara
q
Start
Tara