Reprocessing

2.1

Safety instructions

Reusable products
Reusable products must be reprocessed, otherwise there is an increased risk of
infection.
► Follow the infection prevention policies and reprocessing regulations of the
health-care facility.
► Follow the national infection prevention policies and reprocessing regulations.
► Use validated procedures for reprocessing where applicable.
► Reprocess reusable products after every use.
► Follow the manufacturer's instructions for cleaning agents, disinfectants, and
reprocessing devices.
2.2

Safety information

Electrical hazards
An electrical shock hazard exists when performing cleaning or maintenance
procedures.
► Unplug the power cable (AC or DC) from its power source.
Explosion hazard
A fire and explosion hazard exists when performing cleaning and maintenance
procedures in an oxygen-enriched environment.
► When performing cleaning and maintenance procedures, confirm that the
oxygen supply is turned off and that the equipment is disconnected from the
oxygen supply.
► Switch off or disconnect oxygen supplies during periods of non-use.
► Ensure that oxygen connectors are free from grease and oil.
Equipment damage
Reusing, reprocessing, or sterilizing disposable items can lead to a failure of the
medical device and cause injury to the patient.
► Do not reuse, reprocess, or sterilize disposable items. Disposable items were
designed, tested, and manufactured for one-time use only.
Humidity
Any open gaps in the incubator hood reduce the incubator internal relative humidity.
► Make sure that all hand port gaskets and tubing grommets are properly installed
on the hood.
Faulty products
Signs of wear, e.g., cracks, deformation, discoloration, or peeling, may occur with
reprocessed products.
► Check the products for signs of wear and replace them if necessary..
Supplement to the instructions for use
|
Isolette 8000 plus
Reprocessing