OPERATING INSTRUCTION MANUAL 
 
 
 
INDICATOR/ALARM UNIT RM-592 
 
 
 
 
 
 
 
 
• Request for the Customers 
Read and understand this operating manual before using the detector. 
You must operate the detector in accordance with the operating manual. 
Regardless of warranty period, we shall not make any compensation for accidents and damage 
caused by using this product. 
The compensation shall be made only under the warranty policy of products or parts 
replacement. 
Because this is a safety unit, a regular maintenance for every six months and daily 
maintenance must be performed. 
If you find abnormalities in the detector, please notify them to our local representative 
immediately.  
 
 
 
 
 
 
 
 
 
 
 
 
     
FOR 
 
 
 
PT1E-0971 

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the RM-592 and is the answer not in the manual?

Questions and answers

Summary of Contents for Riken Keiki RM-592

  • Page 1 PT1E-0971                OPERATING INSTRUCTION MANUAL              FOR      INDICATOR/ALARM UNIT RM-592                  • Request for the Customers  • Read and understand this operating manual before using the detector.  • You must operate the detector in accordance with the operating manual.  • Regardless of warranty period, we shall not make any compensation for accidents and damage  caused by using this product.  The compensation shall be made only under the warranty policy of products or parts  replacement.  • Because this is a safety unit, a regular maintenance for every six months and daily ...
  • Page 2: Table Of Contents

        INDEX    1.PRODUCT OUTLINE  1−1.  In the beginning ………………………………………………  3  1−2.  Application for use ……………………………………………  3  1−3.  Identification of each caution marks …………………………  3    2.IMPORTANT INSTRUCTION FOR THE SAFETY  2−1.  Danger items…………………………………………………  4  2−2.  Warning items…………………………………………………  4  2−3.  Caution items…………………………………………………  5    3.PRODUCT FUNCTION  3−1.  External drawing………………………………………………  6  3−2.  Appearance……………………………………………………  6  3−3.  Panel cut-out dimensions……………………………………  7  3−4.  Name of each part and function………………………………  8  3−5.  Block diagram…………………………………………………  9   ...
  • Page 3       6.KINDS OF ALARM AND ITS PERFORMANCE  6−1.  Kinds of alarm…………………………………………………  26  6−2.  Gas alarm   6−2−1.  Gas alarm action……………………………………  26  6−2−2.  Counteraction at gas alarm…………………………  27  6−2−3.  Alarm other than gas detection……………………  27    7.MAINTENANCE CHECK  7−1.  Frequency of maintenance and check items  7−1−1.  Daily check…………………………………………・  28  7−1−2.  Regular maintenance check…………………………  28  7−2.  Maintenance contract for regular check………………………  28  7−3.  Replacement of parts …………………………………………  29  7−4.  Measure when operation is stopped or installation place is moved.  7−4−1.  Stopping the normal operation………………………  29  7−4−2.  Changing the installation place………………………  29  7−5. ...
  • Page 4: 1.Product Outline

        1.PRODUCT OUTLINE    1−1.In  the  beginning    First of all, we wish to express our heartfelt thanks for your purchase of our indicator/alarm unit RM-592.  This instruction manual is just a guide book to operate indicator/alarm unit RM-592 of RM-590 series.  Your kind reading of this manual is requested not only for first user but for already experienced staff.      1−2.Application for use    ○ This is  an exclusive  indicator/alarm unit  for 4-20mA outputs type instrument  and mainly used  as gas  alarm  system in combination with gas detector head to indicate detected gas concentration for environmental control.        1−3.Identification of each caution marks    Indicates an imminently hazardous situation which, if not avoided, will result in death or  DANGER  serious injury.    This signal word is to be limited to the most extreme situation.      Indicates a potentially hazardous situation which, if not avoided, could result in death    WARNING  or serious injury.   ...
  • Page 5: 2.Important Instruction For The Safety

        2.IMPORTANT INSTRUCTION FOR THE SAFETY    2−1.Danger items        DANGER     This is not explosion proof unit.    Do not use it at hazardous area.    It is danger to use this unit under such    environment.        2−2.Warning items        WARNING     ・  Power source    Before supplying the power, confirm that the power source is a designated voltage.        ・  Necessity of protective grounding    Do not cut inside or outside wire for earth, nor disconnect the wire from earth terminal.    In any case, the unit becomes dangerous conditions. ...
  • Page 6: 2−3.Caution Items

      2−3.Caution items        CAUTION       ・.Do not use walky-talky, mobile phone, etc. near this unit.    When  this  unit  receives  the  electric  waves  by  walky-talky  etc  near  from  the  cable,  this  will  effect  the    reading.    When use the walky-talky, operate it at the place where there is no influence of electric wave ejection.        ・  When make power on again, make it after an interval of over 5 seconds.    When make power on within 5 seconds, there may be no normal operation.   ...
  • Page 8: 3−3.Panel Cut-Out Dimensions

      3−3.Panel cut-out dimensions        Dimensions with single case    1. Mounting for 2 steps in vertical and N lines in horizontal                                    +0.6 (36×N−3)   0       2. Mounting for 1 step in vertical and N lines in horizontal                 ...
  • Page 9: 3−4.Name Of Each Part And Function

      3−4.Name of each part and function  *The below pictures show indicator/alarm unit with single case.  ③ ① ⑥ ④ ⑤ ⑦ ⑩ ⑧ ⑨ ⑪ ②   ①LED display・・・・・・・・・・・・・・・・・・・・・・・・ Indicates gas concentration.  ②POWER switch・・・・・・・・・・・・・・・・・・・・・ ON/OFF switch  ③Terminal plates・・・・・・・・・・・・・・・・・・・・・ Connects cables from outside.      For multi-unit case refer to instruction manual for multi-unit case.  ④PW/TR light・・・・・・・・・・・・・・・・・・・・・・・ Combination light for power and trouble indication.    Steady light : Power ON, normal operation    Flashing        : Trouble  ⑤AL1 light・・・・・・・・・・・・・・・・・・・・・・・・・  Flashes at gas alarm (1st alarm)  ⑥AL2 light・・・・・・・・・・・・・・・・・・・・・・・・・  Flashes at gas alarm (2nd alarm)  Alarm lights change continuous lighting after RESET operation.  ⑦SKIP light・・・・・・・・・・・・・・・・・・・・・・・・  Flashes during maintenance mode.  ⑧MODE switch・・・・・・・・・・・・・・・・・・・・・  Uses at maintenance work.  ⑨UP/DOWN switch・・・・・・・・・・・・・・・・・  Uses to select items at maintenance mode, increase or decrease reading  at ...
  • Page 10: 3−5.Block Diagram

      3−5.Block diagram        RM−592    Display part      (PW/TR)  (SKIP)  (AL1)  (AL2)      Power supply part      Operation part  POWER INPUT      (MODE)  (UP)  (DOWN)  (SET)  (DC24V)  Control part        Alarm contact control part      Gas alarm contact(AL1、AL2)    Trouble alarm contact      ...
  • Page 11: 4.How To Use

        4.HOW TO USE    4−1.Before operation  Be sure to keep caution items of use not only for first user but for already experienced staff.  If not keep these caution items, the unit may be defective and correct gas detection may not be performed.      4−2.Mounting/dismounting procedures  Indicator/alarm unit for RM-590 series shall be used after housing it in either exclusive single case or multi-unit  case.    (1)Mounting method          ・Open the front cover of this unit.          ・Confirm that the power switch is at OFF.          ・With the disengagement lever at down position, mount into the case.            (Confirm that the edge connector is put in firmly)          ・Confirm that the disengagement lever is turned upward when indicator/alarm unit is mounted into the case  firmly. ...
  • Page 12: 4−3.Installation Place

      4−3.Installation place      CAUTION       Do not install at places where direct sun drought or where the temperature is drastically changed.        Keep  away  the  system  from  direct  sun  drought  or  where  the  temperature  of  the  system  is  drastically  changed.    If  installed,  inside  the  unit  will  be  built  up  condensation  and  does  not  follow  the  drastic  temperature change. ...
  • Page 13: 4−4.Caution In The System Engineering

      4−4.Caution in the system engineering    CAUTION    Unstable power and noise may cause error of performance and alarm.  For the system to use this unit, it is required to make design based on this manual  descriptions.    (1)  Stable power used  While the system gets stable at power on or instantaneous power failure, the external output and alarm contact  may be on and the care for it must be taken.    In such case, use the standby battery or take an appropriate  action in the receiver side. Supply the following power to this unit      Power voltage  DC24V±10%  Approx 50msec.  Power failure    (For power failure of 50msec, it re-starts)  tolerance time  To  warrant  the  continuous  operation,  install  the  UPS  (Uninterrupted Power Supply) or standby battery outside.    Do  not  share  the  power  involving  with  high  power  load  and  high frequency noise. ...
  • Page 14 SK  contact        V  V     Power supply  Power supply  Relay    (Low voltage relay)    RM−592   −How to think alarm contacts against inductive load−  The spec for alarm contact of RM-592 is described by the conditions of resistance load.  When use the inductive load for alarm contacts, the very high reverse electromotive voltage may be generated and  the following trouble tends to be produced.            ・Contact part of relay is melted adhesively and the contacts can not work.          ・High voltage is put inside of indicator/alarm unit and then, electrical parts may be damaged.          ・As it is big noise, the trouble action may be taken by the reckless drive of CPU.  ・Irrespective of inductive load, there is the possibility of unforeseeable noise instruction for contact.  Above trouble may be generated.  13 ...
  • Page 15: 4−5.Grounding

              CAUTION    ・The inductive load shall not be used in principle (Do not use fluorescent lamp).    ・When use the inductive load, make the contact amplification outside, but the outside relay coil    belong,  use  the  relay  driven  by  the  low  voltage  (within  AC100V)  and  it  is  protected  by  an    appropriate surge killer.        * As the inductive load, there are following samples.        ・Revolving light    * External relay    * Buzzer    * Siren    * Fan    * Fluorescent lamp    *Motor etc.   ...
  • Page 16: 4−6.Wiring

      4−6.Wiring                              (In case of single case)  ① 24V  (+)        TD1 ② DC24V±10%  ③ GND (-)   ④ Note 1  ③  4-20mA    output  3  ① CVVS 1.25mm2, 2-core  ④  instrument  4  ②...
  • Page 17: 5.Operation Method

        5.OPERATION METHOD    5−1.Start-up    ●  B efore turning on the power……………  Before turning on the power ON, keep following caution items.    If not, danger for electric shock or damage to the  unit will be happen.    (1)Make grounding.  (2)Check that wirings with outer units are made correctly.  (3)Check that the supply voltage is within a rating.  (4)Relay contact for outer alarm be activated during adjustments.    Even relay contact works, treat that any  influence will not be given to the outer.  (5)To prevent fire, check that the designated fuse is used.      5−2.Basic performance flow    In normal operation, it is used on detection mode.      Stand-by        Power on        Initial clear (approx. 25sec.)       ...
  • Page 18: 5−3.Start-Up Method

      5−3.Start-up method    5−3−1.Power on      Cover      Power switch      1) Before making power on, check that the unit is connected correctly.  2) The power switch can be seen when the front cover is opened.  3) Put the power switch upward to ON and downward to OFF.  4) Put the power switch on, and power light illuminates.  5) Status of initial clear (about 25 sec.) is displayed as “- - - - “.    ・  Checking of the system (Self-diagnosis)    ・  Output signal: Approx 2.5mA  ・  Gas alarm, trouble alarm (light, contact) shut-off.      17 ...
  • Page 19: 5−4.  Explanation Of Performance (Detection Mode) 5−4−1.  Display Action

      5−4.Explanation of performance (Detection mode)  5−4−1.Display action  There are following two kinds of displays.            (1)LED display :Shows gas concentrations.    (2)LED light  :Shows  power/trouble  (PW/TR),  1st  gas  alarm  (AL1),  2nd  gas  alarm  (AL2)  and  maintenance  (SKIP) condition.    * Contents of indications by LED display and LED light are different from the condition of the unit.      Action of LED light      Preparation □:Light off    ■:Light on    ◎:Flashing     ...
  • Page 20       The  right  「 -0.00 」   is  displayed    when zero level gets down to minus    (-) side at above 10% full scale.  PW/TR    AL2    ■        □    SKIP     AL1    □        □          WARNING    When ...
  • Page 26: 5−6.How To Finish Operation

      5−6.How to finish operation        When finish this operation, turn off power switch of this unit and turn off the main power (DC24V).            WARNING    ・When finish this operation, do it after making point skip with the upper system    (Centralized system).        ・When finish this operation, check the external output and function of outer equipment to be    connected with external alarm contact.    Then, judge whether power can be shut off or not.      When alarm relay contact is energized(option), it is activated when make power (DC24V) off.        25 ...
  • Page 27: 6.Kinds Of Alarm And Its Performance

        6.KINDS OF ALARM AND PERFORMANCE    6−1.Kinds of alarm  There are two kinds of alarms such as gas alarm and trouble alarm.    ・Gas alarm:When detection gas reaches preset alarm level or exceeds it, this starts alarm.  《Latched mode→Self reset after reset operation.》      *  NOTE    Alarm levels are adjusted at 1/3 (1st alarm) and 2/3 (2nd alarm) of full scale.    To prevent wrong operation, this unit provides alarm delayed circuit for 2 seconds.      ・Trouble alarm:Abnormal condition in this unit is detected and gives a trouble alarm.    Except system trouble  「E-00」, it is non-latched mode.    When return to normal from trouble condition, it starts again  from the action of initial clear  after power on see section  「9. TREATMENT AT ABNORMAL CASE」.      6−2.Gas alarm  6−2−1.Gas alarm action  (1)Display  ①Gas concentration    When exceeds detection range (Over scale), LED shows  “∩∩∩∩”.  ②Power light (Power:Green)  It keeps steady light.  ③Alarm light  (AL1:Orange),  (AL2:Red)  There are two alarm levels. ...
  • Page 28: 6−2−2.Counteraction At Gas Alarm

      6−2−2.Counteraction at gas alarm    The counteraction at gas alarm shall follow to the client rule and immediate reaction shall be required.      Generally, the following action is taken.    ①Confirmation of indication value.    *  NOTE    Instantaneous gas leak may get lower at confirmation time.    Except gas alarm, it gets alarm condition temporarily by noise or any other accidental conditions.        ②Based on gas alarm control concentration, it keeps the safety by keeping away people from the monitoring  area.  ③When gas concentration display is continued, close the gas valve and confirm that gas concentration gets  lower enough.  ④Suppose that the leak gas is to remain and provide yourself with protection attire and tool away from danger,  go to the leak site and check the gas residual condition by portable gas detector.  ⑤After checking that there is no danger, the treatment for gas leak shall be taken.    6−2−3.Alarm other than gas detection    t may respond to interference gas  For interference gases, contact the nearest agent or RIKEN KEIKI.  t may be caused by drifting due to the change of long time elapsed sensor  Check the reading by daily check and make calibration through authorized service agent whenever necessary.  t may be caused by the noise generated from spherical machine  Revision of installation location, wirings and the addition etc. of noise measure parts between instrument and    detector shall be required.    This specific measures shall differ from the condition of each site.    It may receive a temporarily noise such as by thunder lightning etc. ...
  • Page 29: 7.Maintenance Check

        7.MAINTENANCE CHECK    This is an important for security and safety.    To maintain the security and enhance the reliability of safety, the  reliability of safety, the regular maintenance check for it shall be absolutely necessary.      7−1.Frequency of maintenance and check items  7−1−1.Daily check    This is the check items carried out by customer.    ①  Check of PW/TR light.  In normal operation, the light is at on condition.  ②  Check of LED reading  Check that LED reading is zero (0).  If it is not “0”, check that gas is free around detector head and make zero adjustment at detector head side.  ③  Alarm test  Check that the alarm light flashes and buzzer sounds.       CAUTION    For the alarm test, refer to  「5-5-6 Alarm test」.        7−1−2.Regular maintenance check    The following items shall be checked at regular check.  ①Daily check  ②Cleaning of this unit  ③Adjustments  ④Function check ...
  • Page 30: 7−3.Replacement Of Parts

      7−3.Replacement of parts          CAUTION    For replacement of sensors and parts etc, contact our service agent or RIKEN KEIKI.          7−4.Measure when operation is stopped or installation place is moved  7−4−1.Stopping the normal operation  Turn off the power switch on the front panel.  Turn off the power (DC24V) at the supply side.    7−4−2.Changing the installation place  When move its place, refer to  「4-2. Installation place」  about its moving place.    Then, for wiring refer to  「4-5. Caution at wiring」.        CAUTION    When move its installation place, be sure to make gas calibration.    The re-adjustment work    including gas calibration shall be requested to contact our service agent or RIKEN KEIKI.      7−5.Storage and treatment when not use for a long time.    This model must be stored under the following environmental conditions.  ・In a dark place under the normal temperature and humidity away from direct sunlight. ...
  • Page 31: 9.Measures At An Abnormal Case

        9.MEASURES AT AN ABNORMAL CASE    9−1.Trouble alarm    Flashing of PW/TR light  ・When PW/TR light is flashing, it is the condition that the following trouble may be caused.  Research its cause and take an appropriate action.  ・The output from the unit at trouble shall be as follows.  4〜20mA output  : Fixed at 0.5mA    Trouble alarm contact  : Relay contact is activated.    LED display  Trouble content  Main cause  E−00  System trouble  Memory trouble in the unit.  Connection trouble of  Detector head disconnection, sensor disconnection, sensor  E−01  detector head  cable disconnection.  Low sample flow  Decrease of sample flow rate in detector head or block of  E−05  rate  sample gas tubing.      9−2.Responding to trouble alarm  When any trouble was found on this unit and detector head, contact our nearest agent or RIKEN KEIKI.      9−3.Before it is thought to be trouble  It may be originated from disconnection and short circuit between units.  ・Check the wirings including peripheral equipment as well as detector head. ...
  • Page 32     (3)PW/TR light flashing  ①  System trouble  「E−00」  <Cause and action>  ・Memory trouble inside of indicator/alarm unit.  Contact our nearest agent or RIKEN KEIKI.  ②  Connection trouble of detector head  「E−01」  <Cause and action>  ・Non-connection of sensor, contact failure of connector and disconnection of sensor cable.  ・Check “whether sensor is connected” or “the sensor cable connector is connected to the sensor.    If can  not be recovered, contact our nearest agent or RIKEN KEIKI.    (4)Fuse disconnection  <Performance>  ・This unit can not work.  <Cause and action>  ・The trouble of this unit or trouble of external power source can be considered.    Search this cause and  after  taking its measures, replace the fuse with the designated one.      *NOTE    The fuse rating for this unit is “250V/1A T”.        31 ...
  • Page 33: 10.Definition Of Word

        10.DEFINITION OF WORD      Calibration  Find relationship of the readings, display values or set-points with the actual values by using the calibration gas.    Maintenance mode  When maintenance is performed on the detector, the alarm contact is disconnected, and a signal to  indicate the maintenance mode status is sent out to the external output signal.    As a result, maintenance  can be performed on a single unit of the detector.  Press MODE switch for 3 seconds, and the maintenance mode starts.    nitial clear  The reading is unstable for seconds after the power is turned on.    To prevent malfunctions for that period, the  alarm contact is deactivated.    In addition, a signal to indicate the initial clear status is sent out to the external  output.    Alarm delay time  A function which temporarily suspends activation to prevent a false alarm caused by noise from its outside.        32 ...
  • Page 34: 11.Product Specifications

    CVVS worth of shield cable・2-core  Various functions  Alarm delay/Peak hold/Alarm trend history  Power supply  DC24V±10%  Power consumption  MAX.5W(MAX.8VA by using Multi case)  Initial clear  Approx.25sec  Operating temperature  0〜40℃(non-rapidly-vary)  Operating humidity  10〜90%RH(non-condensing)  Structure  Housing type for the case  front identification card type(put into Single case or Multi case)  Outer dimension  Approx.36(W)×72(H)×134(D)mm  Weight  Approx.0.10kg(unit only)      11−2.Product composition        Standard compositions  ・Indicator/alarm unit Model RM-592  ・Operating instruction manual…1 vol.  ・Spare fuse…Depends on ordered quantity of the unit.  1〜10 units  1 pce  11〜20 units  2 pcs  21〜50 units  3 pcs  51 units and above  4 pcs    33 ...