Seitron CHEMIST 200 Use And Maintenance Manual page 63

Combustion analyzer
Table of Contents

Advertisement

Instructions for accurate testing
Instructions for accurate testing
Instructions for accurate testing
Instructions for accurate testing
In order to achieve a certain degree of accuracy when conducting flue gas analysis, the following should be respected:
the boiler being checked should be running in steady state conditions
the flue gas analyser should be switched on at least 3 minutes before testing (time to auto-calibrate) with the probe
located in fresh air
the point in which the probe is inserted for analysis has to be at a distance of approximately twice the stack diameter or,
alternatively, as directed by the boiler manufacturer.
the water trap should be completely empty and positioned vertically
before switching off the instrument, extract the probe and wait at least 3 minutes (the CO value has to drop below
10 ppm)
Before returning the instrument to its place, clean the water trap and relative hose; if water is present in the hose clean
the latter by blowing inside.
@IST01380ESE 018350A0 270810
63