Table 204:
The NeuroTrac® product is intended for use in the electromagnetic environment
specified below. The customer or the user of the NeuroTrac® product should assure
that it is used in such an environment.
Immunity
IEC 60601
test
3 Vrms
Conducted RF
150 kHz to 80
IEC 61000-4-6
MHz
Radiated RF
3 V/m
IEC 61000-4-3
80 MHz to 2,5
GHz
NOTE 1 At 80 MHz and 800 MHz, the higher frequency range applies.
NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is af-
fected by absorption and reflection from structures, objects and people.
a Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) tele-
phones and land mobile radios, amateur radio, AM and FM radio broadcast and TV broadcast
cannot be predicted theoretically with accuracy. To assess the electromagnetic environment due
to fixed RF transmitters, an electromagnetic site survey should be considered. If the measured
field strength in the location in which NeuroTrac® product is used exceeds the applicable RF
compliance level above, the NeuroTrac® product should be observed to verify normal operation.
If abnormal performance is observed, additional measures may be necessary, such as reorienting
or relocating the NeuroTrac® product.
b Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 3 V/m.
Guidance and manufacturer's declaration – electromagnetic
Compliance
test level
level
3 Vrms
150 kHz to 80
MHz
3 V/m
80 MHz to 2,5
GH
NeuroTrac
MultiTENS Operation Manual
®
immunity
Electromagnetic environment
Portable and mobile RF communications
equipment should be used no closer to any
part of the
cables, than the recommended separation
distance calculated from the equation ap-
plicable to the frequency of the transmitter.
Recommended separation distance
d = 1.2 √P 150 kHz to 80 MHz,
d = 1.2 √P 80 MHz to 800 MHz
d = 2.3 √P 800 MHz to 2.5GHz
where P is the maximum output power rating
of the transmitter in watts (W) according
to the transmitter manufacturer and d is
the recommended separation distance in
meters (m).
Field strengths from fixed RF transmitters,
as determined by an electromagnetic site
a
survey
, should be less than the compliance
level in each frequency range b .
Interference may occur in the vicinity of
equipment marked with the following
symbol: