Table 202:
The NeuroTrac® product is intended for use in the electromagnetic environment spec-
ified below. The customer or the user of theNeuroTrac® product should assure that it
is used in such an environment, and that precautions regarding that environment are
heeded.
Immunity test
E l e c t ro s t a t i c d i s -
charge (ESD)
IEC 61000-4-2
Po w e r f r e q u e n c y
(50/60 Hz)
magnetic field
IEC 61000-4-8
Guidance and manufacturer's declaration
– electromagnetic immunity
IEC 60601
Compliance
test level
6 kV contact
±6 kV contact
±8 kV air
±8 kV air
3 A/m
3 A/m
NeuroTrac
MultiTENS Operation Manual
®
Electromagnetic environ-
level
Floors should be wood, concrete
or ceramic tile. If floors are cov-
ered with synthetic material, the
relative humidity should be at
least 30 %.
Power frequency magnetic fields
should be at levels characteristic
of a typical location in a typical
commercial or hospital environ-
ment.
29
ment guidance