MOD (modulated TENS mode)
1.
Press B + / - to set the low frequency (Hz, LO) between 2Hz and
200Hz.
Press A + to set the high frequency.
2.
Press B + / - to set the high frequency (Hz, HI) between 2Hz and
200Hz.
Press A + to set the low pulse width.
3.
Press B + / - to set the low pulse width (µS, LO) between 50µS and
450µS.
NOTE: If a frequency of 105 Hz or higher is set, the pulse width
will be restricted to 300µS maximum for safety.
Press A + to set the high pulse width.
4.
Press B + / - to set the high pulse width (µS, HI) between 50µS and
450µS.
NOTE: If a frequency of 105 Hz or higher is set, the pulse width
will be restricted to 300µS maximum for safety.
Press A + to set the milliamp modulation adjustment.
5.
Press B + / - to set a milliamp modulation (ADJ) between 50% and
100% (100% is no milliamp modulation).
Press A + to set the modulation time.
6.
Press B + / - to set the modulation (TME) time between 3 and 60
seconds.
7.
Press the SET button to configure the next phase or press the ESC
button to save the parameters.
BST (BURST TENS mode)
1.
Press B + / - to set the frequency (Hz) between 35Hz and 200Hz
Press A + to set the Pulse Width.
2.
Press B + / - to set the pulse width (µS) between 50µS and 200µS.
Press A+ to set the number of bursts per second.
3.
Press B + / - to set the frequency (FRQ) of bursts per second between
2 and 9.
4.
Press the SET button to configure the next phase or press the ESC
button to save the parameters.
NeuroTrac
MultiTENS Operation Manual
®
11