NMS (Neuromuscular (STIM) mode)
1.
Press B + / - to set the frequency (Hz) between 2Hz and 200Hz
Press A + to set the Pulse Width.
2.
Press B + / - to set the pulse width (µS) between 50µS and 450µS.
NOTE: If frequency 105 Hz or higher is set, the pulse width will
be restricted to 300µS maximum for safety.
3.
Press B + / - to set the work seconds (WRK) between 2 and 99
seconds.
Press A + to set the rest seconds.
4.
Press B + / - to set the rest seconds (RST) between 2 and 99 seconds.
Press A + to set the ramp up time.
6.
Press B + / - to set the ramp up time (R:UP) between 0.0 and 9.9
seconds.
Press A + to set the ramp down time.
7.
Press B + / - to set the ramp down time (R:DN) between 0.0 and 9.9
seconds.
Press A + to set the current type.
8.
Press B + / - to set the current type to either synchronous or
alternating (SYN or ALT).
NOTE: The SYN current stimulate both channels at the same time.
The ALT current stimulates channel A and then channel B.
If SYN is selected, press A + to set the delay between channel A & B.
9.
Press B + / - to set the delay (DLY) between channel A & B from
0.0 to 4.0 seconds.
10.
Press the SET button to configure the next phase or press the ESC
button to save the parameters.
mA Lock
For added safety, the device includes an mA lock function. After the mA is adjusted
and there is no action for 60 seconds, the mA intensity will be locked and pressing the
mA+ button will not have any action. This is to prevent unwanted mA increase. To in-
crease the mA, firstly press the mA- button, which will unlock the mA, and then press
the mA+ button to adjust.
NeuroTrac
MultiTENS Operation Manual
®
12