Intended Use; Prohibited Operating Conditions; Residual Risks - SPECTRO SPECTROLAB Original Operating Instructions

Stationary metal analyzer
Table of Contents

Advertisement

2.3

Intended use

This instrument analyzes and identifies metals. Only such metals may be analyzed
for which the instrument has been calibrated by the manufacturer or by qualified
technical personnel.
2.4

Prohibited operating conditions

Do not operate the instrument if one of the following conditions exists:
If the instrument is faulty or damaged.
If the instrument is not maintained at regular intervals.
If the ambient conditions do not comply with the specification.
If the area where the instrument is used is not a commercial/industrial
premise.
2.5

Residual risks

This instrument has been manufactured in accordance the latest state-of-the-art
technology. Nevertheless, a residual risk cannot be completely excluded.
Caution! Risk of injury and health hazard!
The light inlet window of the optional air optics is located below the spark
stand. When removing the light inlet window, mechanical parts in this area
may cause injuries.
The instrument is operated with argon. Argon is heavier than air. It can
accumulate in a closed room, especially on the floor or lower-lying areas.
SPECTROLAB — 30.07.2015
Protective gloves must be worn.
Ensure to provide adequate ventilation. If necessary, install oxygen
sensors.
Safety
— 11

Hide quick links:

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the SPECTROLAB and is the answer not in the manual?

This manual is also suitable for:

Lavm12Lacm12

Table of Contents

Save PDF