HP 5501A Operating And Service Manual page 239

Laser transducer system
Hide thumbs Also See for 5501A:
Table of Contents

Advertisement

Table 6-4.
English/Metric
Pulse
Output
Electronics
System
Troubleshooting
(Cont'd)
1
CHECKOUT PROCEDURE
*E-Z
mini
test
hook
I
REQUIRED TEST EQUIPMENT
**Test Fixture
-
To fit on front-edge of 10763A English/Metric card for testing purposes. (To be constructed by user).
INSTRUMENT TYPE
Digital Voltmeter
Logic Probe
Oscilloscope
Frequency Counter
"E-Z mini hook7'*
Test Fixture (Front-edge)**
Individual circuit card
operating and service manuals.
Parts:
48-pin
connector
(e.g.
HP
P/N
1251-0335)
Momentary
switch,
SPDT
(e-g.
HP
P/N
3101-1261)
REQUIRED CHARACTERISTICS
Range: -15V to +15V
+
.01V dc
TTL Level
Ability to display sig-
nals between dc and 20 MHz
Totalizing Capability
(e.g. HP 5301, 5328)
Used for test purposes
to activate output cir-
cuitry
Capable of sending
"Start" pulses
The switch sends a "Start" pulse to activate the electronics and initialize the system.
The
switch
should
be
wired
so
that
pins
S&V
are
shorted
when
the
switch
is
pushed.
Pins
R&V
should
be
shorted
when
switch
is
at
rest.
I
Laser Coupler Cards: The minimum number of circuit cards in the coupler are as follows:
(1) 10763A English/Metric
(1) 10760A General Purpose Counter
(1) 10761A Binary Multiplier
(1) 10755A Compensation Interface
(1) 10756A Manual Compensator

Advertisement